EMU40EX BASE UNIT
Report
- Report Number
- 9612330-2020-00012
- Event Type
- Malfunction
- Date Received
- October 29, 2020
- Date of Event
- September 29, 2020
- Report Date
- October 1, 2020
- Manufacturer
- NATUS MEDICAL INCORPORATED
- Product Code
- GWQ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
(B)(6) 2020. REFERENCE (B)(4). THE UNIT AND AFFECTED COMPONENTS HAVE BEEN REQUESTED TO BE RETURNED BACK TO MIDDLETON FOR PRODUCT INVESTIGATION.
21 JANUARY 2021 REF SR# (B)(4). THE CUSTOMER REPORTED THE DEFECTIVE UNIT IS OVERHEATING AND MELTED. 1) NO DEATH OR SERIOUS INJURY CAUSED. 2) NO BURNING SMELL OR EVIDENCE OF BURNING ON THE DEVICE. 3) THE DEVICE WAS BEING USED FOR PATIENT CARE AT THE TIME OF THE EVENT. 4) THE DEVICE DID NOT HAVE CONTACT WITH ANOTHER HOT OBJECT WHICH COULD HAVE CAUSED THE EMU40EX CHASSIS TO MELT. PRODUCT EXAMINATION AND FUNCTIONAL TESTING: INVESTIGATION RESULTS/REPAIR NOTES - STUDY RUNNING FOR 72 HOURS THIS UNIT DID NOT GET HOT DURING THE STUDY RUNNING. IT FELT LITTLE WARM BUT THIS IS NORMAL. NO FAULT FOUND AND NO USER/PATIENT WAS INJURED. DEVICE COMPLIES WITH (B)(4) STANDARDS AND SINCE THE ENCLOSURE IS UL-94 CERTIFIED AND THE PRINTED CIRCUIT BARE BOARD MATERIAL IS FR4 AND QUALIFIED UNDER UL 94. THERE IS NO RISK OF THE DEVICE IGNITING, OR CAUSING HARM TO THE USER/PATIENT. PER HAZARD ID 6.3 OF DOC-010378 EEG/PSG RISK ANALYSIS THE RISK IS LOW - 3. INVESTIGATION RESULT CODE: NEURO SBU/NO ISSUES NOTED.
CUSTOMER REPORTED PART OVER HEATED AND WAS HOT TO TOUCH. NO PATIENT INJURY.
CUSTOMER REPORTED PART OVER HEATED AND WAS HOT TO TOUCH.
(B)(6) 2020 REFERENCE SR#1-740234502. ADDITIONAL INFORMATION HAS BEEN REQUESTED FROM THE CUSTOMER.
CUSTOMER REPORTED PART OVER HEATED AND WAS HOT TO TOUCH.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1224661 | EMU40EX BASE UNIT | EMU40EX BASE UNIT | GWQ | NATUS MEDICAL INCORPORATED | 006550 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |