FDA Adverse Event Malfunction Summary report: N

EMU40EX BASE UNIT

MDR report key: 10756947 · Received October 29, 2020

Report

Report Number
9612330-2020-00012
Event Type
Malfunction
Date Received
October 29, 2020
Date of Event
September 29, 2020
Report Date
October 1, 2020
Manufacturer
NATUS MEDICAL INCORPORATED
Product Code
GWQ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

(B)(6) 2020. REFERENCE (B)(4). THE UNIT AND AFFECTED COMPONENTS HAVE BEEN REQUESTED TO BE RETURNED BACK TO MIDDLETON FOR PRODUCT INVESTIGATION.

Additional Manufacturer Narrative · 0

21 JANUARY 2021 REF SR# (B)(4). THE CUSTOMER REPORTED THE DEFECTIVE UNIT IS OVERHEATING AND MELTED. 1) NO DEATH OR SERIOUS INJURY CAUSED. 2) NO BURNING SMELL OR EVIDENCE OF BURNING ON THE DEVICE. 3) THE DEVICE WAS BEING USED FOR PATIENT CARE AT THE TIME OF THE EVENT. 4) THE DEVICE DID NOT HAVE CONTACT WITH ANOTHER HOT OBJECT WHICH COULD HAVE CAUSED THE EMU40EX CHASSIS TO MELT. PRODUCT EXAMINATION AND FUNCTIONAL TESTING: INVESTIGATION RESULTS/REPAIR NOTES - STUDY RUNNING FOR 72 HOURS THIS UNIT DID NOT GET HOT DURING THE STUDY RUNNING. IT FELT LITTLE WARM BUT THIS IS NORMAL. NO FAULT FOUND AND NO USER/PATIENT WAS INJURED. DEVICE COMPLIES WITH (B)(4) STANDARDS AND SINCE THE ENCLOSURE IS UL-94 CERTIFIED AND THE PRINTED CIRCUIT BARE BOARD MATERIAL IS FR4 AND QUALIFIED UNDER UL 94. THERE IS NO RISK OF THE DEVICE IGNITING, OR CAUSING HARM TO THE USER/PATIENT. PER HAZARD ID 6.3 OF DOC-010378 EEG/PSG RISK ANALYSIS THE RISK IS LOW - 3. INVESTIGATION RESULT CODE: NEURO SBU/NO ISSUES NOTED.

Description of Event or Problem · 0

CUSTOMER REPORTED PART OVER HEATED AND WAS HOT TO TOUCH. NO PATIENT INJURY.

Description of Event or Problem · 0

CUSTOMER REPORTED PART OVER HEATED AND WAS HOT TO TOUCH.

Additional Manufacturer Narrative · 1

(B)(6) 2020 REFERENCE SR#1-740234502. ADDITIONAL INFORMATION HAS BEEN REQUESTED FROM THE CUSTOMER.

Description of Event or Problem · 1

CUSTOMER REPORTED PART OVER HEATED AND WAS HOT TO TOUCH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1224661 EMU40EX BASE UNIT EMU40EX BASE UNIT GWQ NATUS MEDICAL INCORPORATED 006550

Patients

Seq Age Sex Outcome Treatment
1