FDA Adverse Event
Malfunction
Summary report: N
HEATER-COOLER SYSTEM 3T
MDR report key: 10755735
·
Received October 29, 2020
Report
- Report Number
- 9611109-2020-00601
- Event Type
- Malfunction
- Date Received
- October 29, 2020
- Date of Event
- September 29, 2020
- Report Date
- October 26, 2020
- Manufacturer
- LIVANOVA DEUTSCHLAND
- Product Code
- DWC
- PMA / PMN Number
- K191402
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THERE WAS NO PATIENT INVOLVEMENT. LIVANOVA (B)(4) MANUFACTURES THE HEATER-COOLER SYSTEM 3T . THE INCIDENT OCCURRED IN (B)(6). THE TECHNICIAN IN CHARGE CONFIRMED THE REPORTED ISSUE AND TRACED IT BACK TO THE PATIENT PUMP. THE PART WAS REPLACED AND THE ISSUE SOLVED. SUBSEQUENT FUNCTIONAL VERIFICATION TESTING WAS COMPLETED WITHOUT FURTHER ISSUES AND THE UNIT WAS RETURNED TO SERVICE. THE MOST LIKELY ROOT CAUSE OF THE REPORTED EVENT IS BEARING CORROSION FORMATION DUE TO ENVIRONMENTAL CONDITION IN WHICH THE PUMP WORKED.
Description of Event or Problem · 1
LIVANOVA (B)(4) RECEIVED A REPORT THAT A HEATER-COOLER SYSTEM 3T DISPLAYED AN ERROR CODE ASSOCIATED TO PATIENT PUMPS DURING SETUP. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1223827 | HEATER-COOLER SYSTEM 3T | CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS | DWC | LIVANOVA DEUTSCHLAND | 16-02-85 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |