FDA Adverse Event Malfunction Summary report: N

HEATER-COOLER SYSTEM 3T

MDR report key: 10755735 · Received October 29, 2020

Report

Report Number
9611109-2020-00601
Event Type
Malfunction
Date Received
October 29, 2020
Date of Event
September 29, 2020
Report Date
October 26, 2020
Manufacturer
LIVANOVA DEUTSCHLAND
Product Code
DWC
PMA / PMN Number
K191402
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THERE WAS NO PATIENT INVOLVEMENT. LIVANOVA (B)(4) MANUFACTURES THE HEATER-COOLER SYSTEM 3T . THE INCIDENT OCCURRED IN (B)(6). THE TECHNICIAN IN CHARGE CONFIRMED THE REPORTED ISSUE AND TRACED IT BACK TO THE PATIENT PUMP. THE PART WAS REPLACED AND THE ISSUE SOLVED. SUBSEQUENT FUNCTIONAL VERIFICATION TESTING WAS COMPLETED WITHOUT FURTHER ISSUES AND THE UNIT WAS RETURNED TO SERVICE. THE MOST LIKELY ROOT CAUSE OF THE REPORTED EVENT IS BEARING CORROSION FORMATION DUE TO ENVIRONMENTAL CONDITION IN WHICH THE PUMP WORKED.

Description of Event or Problem · 1

LIVANOVA (B)(4) RECEIVED A REPORT THAT A HEATER-COOLER SYSTEM 3T DISPLAYED AN ERROR CODE ASSOCIATED TO PATIENT PUMPS DURING SETUP. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1223827 HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS DWC LIVANOVA DEUTSCHLAND 16-02-85

Patients

Seq Age Sex Outcome Treatment
1