BD RAPID DETECTION OF SARS-COV-2 VERITOR
Report
- Report Number
- 1119779-2020-00825
- Event Type
- Malfunction
- Date Received
- October 29, 2020
- Date of Event
- October 8, 2020
- Report Date
- April 28, 2021
- Manufacturer
- BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD.
- Product Code
- QKP
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4). INVESTIGATION SUMMARY: BD HAS RECEIVED SEVERAL CUSTOMER COMPLAINTS FOR FALSE POSITIVE RESULTS, WHEN USING BD SARS-COV-2 REAGENTS FOR BD VERITOR¿ SYSTEM. THE CURRENT INVESTIGATION CONCERNS MULTIPLE LOTS OF BD SARS-COV-2 REAGENTS FOR BD VERITOR¿ SYSTEM. BD TAKES A SYSTEMATIC APPROACH TO INVESTIGATING FALSE POSITIVE COMPLAINTS THAT ARE RECEIVED. THIS APPROACH INVOLVES REVIEW OF MANUFACTURING BATCH HISTORY RECORDS, TESTING OF RETENTION SAMPLES, AND TESTING OF CUSTOMER RETURNED SAMPLES IF APPLICABLE. RETAIN SAMPLES WERE TESTED FOR BATCH #: 0221748. THE DEFECT COULD NOT BE REPLICATED AND THEREFORE THE COMPLAINT COULD NOT BE CONFIRMED. A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 0221748. THE INVESTIGATION DID NOT FIND A ROOT CAUSE FOR THE FALSE POSITIVE RESULTS THAT WERE OBSERVED. IT IS RECOMMENDED THAT EACH CUSTOMER REVIEW THEIR WORKFLOW CAREFULLY TO ENSURE THAT THE PACKAGE INSERT IS BEING FOLLOWED AS WRITTEN. THE ROOT CAUSE IS UNDER INVESTIGATION AND WILL BE DOCUMENTED IN OUR QUALITY SYSTEM. BD CANNOT CONFIRM THE COMPLAINT BASED ON THE INVESTIGATION THAT WAS PERFORMED. A CORRECTIVE AND PREVENTIVE ACTION (CAPA#1878253) IS ALREADY INITIATED TO INVESTIGATE THE ROOT CAUSE AND SOME MITIGATION ACTIONS ARE ALREADY BEING ADDRESSED.
THE FOLLOWING FIELDS HAVE BEEN UPDATED WITH ADDITIONAL INFORMATION: B.5. DESCRIBE EVENT OR PROBLEM: IT WAS REPORTED WHILE TESTING FOR SARS COV-2 2 FALSE POSITIVE RESULTS WERE OBTAINED. CONFIRMATORY TESTING WAS PERFORMED USING PCR TEST METHOD AND THE RESULTS WERE NEGATIVE. THE CUSTOMER STATED THE RESIDENTS TESTED WERE ASYMPTOMATIC. THIS TEST IS NOT INTENDED FOR USE ON ASYMPTOMATIC PATIENTS AND WAS THEREFORE USED OFF LABEL. THERE WAS NO REPORT OF PATIENT IMPACT. (B)(4). D.3. MEDICAL DEVICE MANUFACTURER: BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD. ¿ SUZHOU, CHINA / 215021 D.4. MEDICAL DEVICE LOT #: 0221748 D.4. MEDICAL DEVICE EXPIRATION DATE: 2021-01-21 G.2 MANUFACTURING LOCATION: BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD. ¿ SUZHOU, CHINA / 215021 H.4. DEVICE MANUFACTURE DATE: 2020-08-08.
IT WAS REPORTED WHILE TESTING FOR SARS COV-2 A POTENTIAL FALSE POSITIVE RESULT WAS OBTAINED. THERE WAS NO FORM OF REPEAT OR CONFIRMATORY TESTING REPORTED BY THE CUSTOMER. BD HAS ATTEMPTED TO CONTACT THE CUSTOMER MULTIPLE TIMES FOR ADDITIONAL INFORMATION REGARDING THIS EVENT WITH NO RESPONSE FROM THE CUSTOMER AT THIS TIME. THERE WAS NO REPORT OF PATIENT IMPACT. (B)(4).
IT WAS REPORTED WHILE TESTING FOR SARS COV-2 2 FALSE POSITIVE RESULTS WERE OBTAINED. CONFIRMATORY TESTING WAS PERFORMED USING PCR TEST METHOD AND THE RESULTS WERE NEGATIVE. THE CUSTOMER STATED THE RESIDENTS TESTED WERE ASYMPTOMATIC. THIS TEST IS NOT INTENDED FOR USE ON ASYMPTOMATIC PATIENTS AND WAS THEREFORE USED OFF LABEL. THERE WAS NO REPORT OF PATIENT IMPACT. (B)(4).
MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. DEVICE MANUFACTURE DATE: UNKNOWN. (B)(4).
IT WAS REPORTED WHILE TESTING FOR SARS COV-2 A POTENTIAL FALSE POSITIVE RESULT WAS OBTAINED. THERE WAS NO FORM OF REPEAT OR CONFIRMATORY TESTING REPORTED BY THE CUSTOMER. BD HAS ATTEMPTED TO CONTACT THE CUSTOMER MULTIPLE TIMES FOR ADDITIONAL INFORMATION REGARDING THIS EVENT WITH NO RESPONSE FROM THE CUSTOMER AT THIS TIME. THERE WAS NO REPORT OF PATIENT IMPACT. (B)(4).