FDA Adverse Event
Malfunction
Summary report: N
SECA 403 BABY SCALE CART
MDR report key: 10754719
·
Received October 29, 2020
Report
- Report Number
- 10754719
- Event Type
- Malfunction
- Date Received
- October 29, 2020
- Date of Event
- August 24, 2020
- Report Date
- October 21, 2020
- Manufacturer
- SECA
- Product Code
- FRW
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- FL
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Description of Event or Problem · 1
A SECA BABY CART MODEL 403 WAS DROPPED OFF IN THE CLINICAL ENGINEERING DEPARTMENT WITH ONE OF ITS REAR WHEELS DETACHED. AFTER INSPECTION, IT APPEARS THE REAR WHEEL'S SLOTTED SOCKET THAT HELD THE WHEEL'S BOLT HAD EXPANDED AND WORN. THIS BABY CART'S FRONT WHEELS HAD ALREADY BEEN REMEDIATED FOR A CLASS 2 DEVICE RECALL WHERE THE FRONT BRAKE WHEELS COULD LOOSEN AND DETACH. RECALL NUMBER Z-1429-2020. MANUFACTURER RESPONSE FOR BABY SCALE CART, SECA (PER SITE REPORTER) THEY ACKNOWLEDGE AND THANK US FOR SHARING. NO FURTHER COMMUNICATIONS ABOUT THIS ISSUE WITH THE BABY CART RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1221381 | SECA 403 BABY SCALE CART | SCALE, PATIENT | FRW | SECA | 403 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |