FDA Adverse Event Malfunction Summary report: N

SECA 403 BABY SCALE CART

MDR report key: 10754719 · Received October 29, 2020

Report

Report Number
10754719
Event Type
Malfunction
Date Received
October 29, 2020
Date of Event
August 24, 2020
Report Date
October 21, 2020
Manufacturer
SECA
Product Code
FRW
Product Problem
Yes
Report Source
User Facility report
Reporter Location
FL
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Description of Event or Problem · 1

A SECA BABY CART MODEL 403 WAS DROPPED OFF IN THE CLINICAL ENGINEERING DEPARTMENT WITH ONE OF ITS REAR WHEELS DETACHED. AFTER INSPECTION, IT APPEARS THE REAR WHEEL'S SLOTTED SOCKET THAT HELD THE WHEEL'S BOLT HAD EXPANDED AND WORN. THIS BABY CART'S FRONT WHEELS HAD ALREADY BEEN REMEDIATED FOR A CLASS 2 DEVICE RECALL WHERE THE FRONT BRAKE WHEELS COULD LOOSEN AND DETACH. RECALL NUMBER Z-1429-2020. MANUFACTURER RESPONSE FOR BABY SCALE CART, SECA (PER SITE REPORTER) THEY ACKNOWLEDGE AND THANK US FOR SHARING. NO FURTHER COMMUNICATIONS ABOUT THIS ISSUE WITH THE BABY CART RECEIVED.