FDA Adverse Event Malfunction Summary report: N

BD INTIMA-II CLOSED IV CATHETER SYSTEM

MDR report key: 10752118 · Received October 28, 2020

Report

Report Number
3006948883-2020-00633
Event Type
Malfunction
Date Received
October 28, 2020
Date of Event
September 20, 2020
Report Date
November 23, 2020
Manufacturer
BD (SUZHOU)
Product Code
FOZ
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 7234030. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING; IN LIEU OF THE AFFECTED DEVICE, FUNCTIONAL TESTING WAS PERFORMED ON RETENTION SAMPLES FOR THIS LOT, THE RESULTS OF THESE SHOW THAT THE TESTED UNITS PERFORMED WITHIN PRODUCT SPECIFICATIONS. UNFORTUNATELY WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM EXPERIENCED LEAKAGE DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WHEN USING THE INDWELLING NEEDLE TO PUNCTURE THE PATIENT, IT WAS FOUND THAT THE INDWELLING NEEDLE HANDLE AND CATHETER TUBING WERE LEAKING.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM EXPERIENCED LEAKAGE DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WHEN USING THE INDWELLING NEEDLE TO PUNCTURE THE PATIENT, IT WAS FOUND THAT THE INDWELLING NEEDLE HANDLE AND CATHETER TUBING WERE LEAKING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1217735 BD INTIMA-II CLOSED IV CATHETER SYSTEM INTERVASCULAR CATHETER FOZ BD (SUZHOU) 7234030

Patients

Seq Age Sex Outcome Treatment
1