FDA Adverse Event
Malfunction
Summary report: N
GLOOKO WEB APPLICATION
MDR report key: 10752087
·
Received October 28, 2020
Report
- Report Number
- 3009137498-2020-00009
- Event Type
- Malfunction
- Date Received
- October 28, 2020
- Date of Event
- September 30, 2020
- Report Date
- October 28, 2020
- Manufacturer
- GLOOKO INC
- Product Code
- OUG
- PMA / PMN Number
- N/A#S-P#N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
A HCP USER REPORTED THAT CERTAIN SETTINGS FROM THE INSULET OMNIPOD DASH SYSTEM, THAT ARE UPLOADED TO THE GLOOKO DEVICE SYSTEM VIA AN INSULET CLOUD TO GLOOKO CLOUD CONNECTION, WERE NOT GETTING DISPLAYED ON THE GLOOKO WEB APPLICATION. NO ADVERSE EVENTS WERE REPORTED. THE GLOOKO ENGINEERING TEAM WAS ABLE TO REPRODUCE THE ISSUE AND A SOFTWARE CHANGE WAS IMPLEMENTED. PRIOR TO THE GLOOKO SOFTWARE CHANGE, A WORKAROUND WAS PROVIDED TO THE AFFECTED USERS VIA A MANUAL UPLOAD USING INSULET'S USB CABLE INSTEAD OF THE CLOUD TO CLOUD CONNECTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1216059 | GLOOKO WEB APPLICATION | GLOOKO WEB APPLICATION | OUG | GLOOKO INC | 20.5.1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |