FDA Adverse Event Malfunction Summary report: N

OMNIPOD INSULIN MANAGEMENT SYSTEM

MDR report key: 10749925 · Received October 28, 2020

Report

Report Number
3004464228-2020-16940
Event Type
Malfunction
Date Received
October 28, 2020
Date of Event
October 24, 2020
Report Date
October 26, 2020
Manufacturer
INSULET CORPORATION
Product Code
LZG
PMA / PMN Number
K162296
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

NO ISSUES WERE FOUND THAT WOULD RESULT IN A NEEDLE MECHANISM FAILURE. THE POD WAS IN POD STATE 14 INDICATING THAT POD ACTIVATION WAS NOT COMPLETED. INITIAL ATTEMPTS TO DOWNLOAD THE POD WERE UNSUCCESSFUL. AN EXTERNAL POWER SOURCE WAS REQUIRED TO RETRIEVE THE POD DATA. MEASUREMENT OF THE BATTERY VOLTAGES FOUND THE BOTTOM AND MIDDLE BATTERIES TO BE DEPLETED. IT COULD NOT BE DETERMINED WHEN THE BATTERIES BECAME DEPLETED. PER NEW INFORMATION, THE FOLLOWING WERE UPDATED. D1: BRAND NAME CHANGED FROM OMNIPOD INSULIN PUMP TO OMNIPOD INSULIN MANAGEMENT SYSTEM. D4: MODEL NO CHANGED FROM 14810 TO 19191. D4: LOT NO CHANGED FROM BLANK TO L45884. D4: CATALOG NO CHANGED FROM ZXY425 TO ZXP425. D4: EXPIRATION DATE CHANGED FROM BLANK TO 12/29/2021. D4: UNIQUE IDENTIFIER (UDI) # CHANGED FROM (B)(4). G5: PMA/510(K) # CHANGED FROM K192659 TO K162296. H4: DEVICE MFG DATE CHANGED FROM BLANK TO 6/29/2020.

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT BEEN RETURNED/RECEIVED TO DATE. IF THE DEVICE IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED WITH THE INVESTIGATION RESULTS. WE ARE UNABLE TO CONFIRM THE REPORTED NEEDLE MECHANISM FAILURE OR TO DETERMINE ITS ROOT CAUSE. NO LOT RELEASE RECORDS WERE REVIEWED, AS THE PRODUCT LOT NUMBER WAS NOT PROVIDED.

Description of Event or Problem · 1

IT WAS REPORTED THE NEEDLE DID NOT DEPLOY, INDICATING A NEEDLE MECHANISM FAILURE. THE PATIENT'S BLOOD GLUCOSE LEVELS WERE UNAFFECTED, AND THE POD WAS NOT WORN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1217417 OMNIPOD INSULIN MANAGEMENT SYSTEM PUMP, INFUSION, INSULIN LZG INSULET CORPORATION 19191 L45884

Patients

Seq Age Sex Outcome Treatment
1