Narratives
INVESTIGATION, ROOT CAUSE, AND CORRECTION. ROOT CAUSE: IN DIVA SOFTWARE (B) (4) OR (B) (4) THE STATISTICS VIEW UPDATE IS DONE USING DIFFERENT THREADS AND ON THE XW4400 SYSTEM BECAUSE THE SYSTEM USES TWO PROCESSORS. THE UPDATES TO THE VIEW BY THESE DIFFERENT THREADS CAN GET INTO A CONDITION WHERE INTERMITTENTLY THE VIEW MIGHT NOT GET THE COMPLETE UPDATE AND HENCE THE STATISTICS VIEW VALUES CAN BE DIFFERENT FROM THE POPULATION HIERARCHY. WORK AROUND: THE FREQUENCY OF THE PROBLEM WILL BE REDUCED, BUT NOT PREVENTED, BY HAVING ONE OR MORE PARAMETER STATISTICS (IE, MEAN, MEDIAN, PERCENT CV, AND SD) DISPLAYED IN EACH STATISTICS VIEW. IN ORDER TO DETERMINE IF THE STATISTICS HAVE BEEN IMPACTED, USER WILL BE INSTRUCTED TO RELOAD DATA BY CLOSING AND REOPENING AN EXPERIMENT BEFORE REPORTING RESULTS. CORRECTION: - BD WILL SEND CUSTOMER LETTER AND PROVIDED A WORK AROUND TO THE USERS. - BD BIOSCIENCES WILL RELEASE A VERSION OF BD FACSDIVA SOFTWARE TO CORRECT THIS PROBLEM. THIS INFORMATION CONSTITUTES THE INITIAL AND FINAL REPORT.
THIS PROBLEM CAN OCCUR ON BD FACSDIVA (B) (4) AND (B) (4) WHEN COMBINED WITH A SPECIFIC COMPUTER SYSTEM CONFIGURATION. THIS PROBLEM HAS NOT BEEN REPRODUCED IN OTHER COMPUTER SYSTEMS OR IN OTHER BD FACSDIVA SOFTWARE VERSIONS DISTRIBUTED BY (B) (4). THIS ISSUE WAS DISCOVERED BY A CUSTOMER. OUR INVESTIGATION DETERMINED THAT THIS ERROR OCCURS WITHIN THE FUNCTION THAT UPDATES THE RESULTS STATISTICS IN BD FACSDIVA SOFTWARE (B) (4) AND (B) (4) RUNNING ON THE (B) (6) COMPUTER SYSTEM. WHEN THE GATING IS ADJUSTED, THE STATISTICS ASSOCIATED WITH THE UPDATED GATES MAY NOT BE UPDATED. THIS ERROR OCCURS WHEN GATES OR COMPENSATION ARE ADJUSTED. NOTE: PREVIOUSLY SUBMITTED IN A BUNDLED REPORT (MFR REPORT#: 2916837-2007-00005).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BD FACSDIVA VERSION 6.1 | BD FACSDIVA | GKZ | BD BIOSCIENCES | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |