BD VACUTAINER® BUFFERED SODIUM CITRATE (9NC) BLOOD COLLECTION TUBES

MDR report key: 10742410 · Received October 27, 2020

Report

Report Number
1917413-2020-01019
Event Type
Malfunction
Date Received
October 27, 2020
Date of Event
October 6, 2020
Report Date
October 29, 2020
Manufacturer
BECTON, DICKINSON & CO. (BROKEN BOW)
Product Code
JKA
UDI-DI
50382903630832
PMA / PMN Number
K013971
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY BD HAD NOT RECEIVED SAMPLES OR PHOTOS FROM THE CUSTOMER FOR INVESTIGATION. THEREFORE, 10 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY FUNCTIONAL TESTING AND NO ISSUES WERE OBSERVED RELATING TO OVERFILL AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING USE WITH A BD VACUTAINER® BUFFERED SODIUM CITRATE (9NC) BLOOD COLLECTION TUBES THE TUBES WERE OVERFILLING. THIS OCCURRED ON 10 SEPARATE OCCASIONS WITH BOTH LOTS, HOWEVER, THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: (1 OF 4 COMPLAINTS) IT IS REPORTED THAT CUSTOMER IS EXPERIENCING OVER FILLING.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. MULTIPLE LOT NUMBERS: THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 0133233. MEDICAL DEVICE EXPIRATION DATE: 021-02-28. DEVICE MANUFACTURE DATE: 2020-05-12. MEDICAL DEVICE LOT #: 0167176. MEDICAL DEVICE EXPIRATION DATE: 2021-03-31. DEVICE MANUFACTURE DATE: 2020-06-15. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING USE WITH A BD VACUTAINER® BUFFERED SODIUM CITRATE (9NC) BLOOD COLLECTION TUBES THE TUBES WERE OVERFILLING. THIS OCCURRED ON 10 SEPARATE OCCASIONS WITH BOTH LOTS, HOWEVER, THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: (1 OF 4 COMPLAINTS). IT IS REPORTED THAT CUSTOMER IS EXPERIENCING OVER FILLING.