FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER BUFFERED SODIUM CITRATE (9NC) BLOOD COLLECTION TUBES

MDR report key: 10742382 · Received October 27, 2020

Report

Report Number
1917413-2020-01023
Event Type
Malfunction
Date Received
October 27, 2020
Date of Event
October 2, 2020
Report Date
November 5, 2020
Manufacturer
BECTON, DICKINSON & CO. (BROKEN BOW)
Product Code
JKA
UDI-DI
50382903630832
PMA / PMN Number
K013971
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: BD HAD NOT RECEIVED SAMPLES OR PHOTOS FROM THE CUSTOMER FOR INVESTIGATION. THEREFORE, 10 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY FUNCTIONAL TESTING] AND NO ISSUES WERE OBSERVED RELATING TO OVERFILL AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING USE WITH A BD VACUTAINER® BUFFERED SODIUM CITRATE (9NC) BLOOD COLLECTION TUBES THE TUBE OVER FILLED. THIS OCCURRED ON 10 OCCASIONS WITH BOTH LOTS, HOWEVER, THERE WAS NO REPORTED PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: (3 OF 4 COMPLAINTS) IT IS REPORTED THAT CUSTOMER IS EXPERIENCING OVER FILLING.

Additional Manufacturer Narrative · 1

MULTIPLE LOT NUMBERS: THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 0133233. MEDICAL DEVICE EXPIRATION DATE: 2021-02-28. DEVICE MANUFACTURE DATE: 020-05-12. MEDICAL DEVICE LOT #: 0167176. MEDICAL DEVICE EXPIRATION DATE: 2021-03-31. DEVICE MANUFACTURE DATE: 2020-06-15. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING USE WITH A BD VACUTAINER® BUFFERED SODIUM CITRATE (9NC) BLOOD COLLECTION TUBES THE TUBE OVER FILLED. THIS OCCURRED ON 10 OCCASIONS WITH BOTH LOTS, HOWEVER, THERE WAS NO REPORTED PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: (3 OF 4 COMPLAINTS). IT IS REPORTED THAT CUSTOMER IS EXPERIENCING OVER FILLING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1209732 BD VACUTAINER BUFFERED SODIUM CITRATE (9NC) BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO. (BROKEN BOW) 363083 SEE H.10 50382903630832

Patients

Seq Age Sex Outcome Treatment
1