FDA Adverse Event Other Summary report: N

SOMATOM EMOTION 16

MDR report key: 1074019 · Received July 9, 2008

Report

Report Number
2240869-2008-00007
Event Type
Other
Date Received
July 9, 2008
Date of Event
June 6, 2008
Report Date
June 9, 2008
Manufacturer
SIEMENS SHANGHAI MEDICAL EQUIPMENT, LTD.
Product Code
JAK
PMA / PMN Number
K050297
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS BALANCE BLOCK IS SECURED TO THE GANTRY DISC BY 3 BOLTS AND NUTS, THE ASSEMBLY PROCESS IS CLEARLY DEFINED IN THE MANUFACTURING INSTRUCTION. DUE TO A MISTAKE MADE DURING PRODUCTION, THE NUTS USED TO SECURE THE WEIGHT WERE NOT TIGHTEN TO THE RECOMMENDED SPECS CAUSING THE BALANCE BLOCK TO LOOSEN AND DETACH FROM THE GANTRY. ACCORDING TO THE INVESTIGATION PERFORMED BY OUR INTN'L AFFILIATE, WE HAVE NO OTHER REPORTED EVENT, WE REPORT THIS AS A SINGLE CASE. NOTE: THERE WAS NO REPORTED INJURY DURING THIS EVENT. WE WERE UNABLE TO DESELECT "FEMALE" IN PT INFO.

Description of Event or Problem · 1

IT WAS REPORTED WHILE DURING DAILY CALIBRATION BEFORE PT SCANS, ONE OF THE BALANCE BLOCKS AFIXED TO THE ROTATING GANTRY DISC, DETACHED FROM THE GANTRY THROUGH THE GAP BETWEEN FRONT AND REAR COVERS. THE BALANCE BLOCK HIT THE EXTENSION OF PT CABLE AND FELL TO FLOOR. SINCE THIS EVENT OCCURRED DURING THE DAILY CALIBRATION, NO PT EXAMINATION WERE BEING CONDUCTED, AND NO INJURIES WERE REPORTED. DAMAGE TO THE SCANNER WAS PRIMARILY LIMITED TO INTERNAL COMPONENTS WITH SOME COSMETIC DAMAGE TO THE PT TABLE AND THE GANTRY COVERS. ALL DAMAGED PARTS WERE REPLACED, AND THE GANTRY WAS REBALANCED BY LOCAL SERVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SOMATOM EMOTION 16 SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED JAK SIEMENS SHANGHAI MEDICAL EQUIPMENT, LTD. 10165977 NA

Patients

Seq Age Sex Outcome Treatment
1