FDA Adverse Event Malfunction Summary report: N

MCLAREN PROTON THERAPY SYSTEM

MDR report key: 10739687 · Received October 27, 2020

Report

Report Number
3015178562-2020-00002
Event Type
Malfunction
Date Received
October 27, 2020
Date of Event
September 17, 2020
Report Date
October 26, 2020
Manufacturer
MCLAREN HEALTH CARE
Product Code
LHN
PMA / PMN Number
K160063
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
RISK MANAGER

Narratives

Additional Manufacturer Narrative · 1

THE CBCT BEGAN ROTATING AND ACQUIRED IMAGES FOR THE PATIENT ALIGNMENT PROCEDURE. DURING THE IMAGE ACQUISITION, THERE WAS AN INTERRUPTION OF THE HEARTBEAT THAT TRIGGERED THE SAFETY SYSTEM IN XPS. THIS RESULTED IN SETTING THE MOTORS TO ZERO TORQUE, AND WITHIN 1.6 SECONDS, ENGAGING THE BRAKE TO STOP THE ROTATION OF CBCT. UNFORTUNATELY, THE BRAKES DID NOT ENGAGE FAST ENOUGH TO STOP THE XPS FROM ROTATING. TO PREVENT THIS FROM HAPPENING AGAIN, THE ENGINEER HAS REDUCED THE DELAY TIME FROM 1600 MSEC TO 100 MSEC FOR THE BRAKING SYSTEM TO GET ACTIVATED AFTER THE SAFETY SYSTEM IS TRIGGERED. FURTHER TESTING HAS NOT SHOWN ANY FAILURE, AND THE ENGINEERING TEAM CONCLUDED THAT WITH THIS CHANGE, THE SYSTEM IS SAFE TO OPERATE.

Description of Event or Problem · 1

DURING PATIENT TREATMENT IN TREATMENT ROOM 1 THE IMAGING C-RING LOST TORQUE AROUND THE PATIENT ON THE TREATMENT COUCH, CAUSING THE X-RAY TUBE TO APPEAR TO COLLIDE WITH THE NOZZLE. THERE WERE NO ERROR WARNINGS, OR OTHER NOTIFICATION OF DEVICE MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1205331 MCLAREN PROTON THERAPY SYSTEM PROTON THERAPY SYSTEM LHN MCLAREN HEALTH CARE N/A N/A

Patients

Seq Age Sex Outcome Treatment
1 65 YR MCLAREN PROTON THERAPY SYSTEM