FDA Adverse Event Malfunction Summary report: N

HEATER-COOLER SYSTEM 3T

MDR report key: 10738904 · Received October 27, 2020

Report

Report Number
9611109-2020-00598
Event Type
Malfunction
Date Received
October 27, 2020
Date of Event
September 28, 2020
Report Date
October 26, 2020
Manufacturer
LIVANOVA DEUTSCHLAND
Product Code
DWC
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THERE WAS NO PATIENT INVOLVEMENT. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA, AND IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K191402). LIVANOVA (B)(6) MANUFACTURES THE HEATER-COOLER SYSTEM 3T . THE INCIDENT OCCURRED IN (B)(6). THE ROOT CAUSE OF THE REPORTED LEAK WAS CAUSED BY AN ASSEMBLY ERROR. THE YELLOW PLUG ON THE VALVE BLOCK OF THE CARDIOPLEGIA BRIDGE WAS MISSING.

Description of Event or Problem · 1

LIVANOVA (B)(6) RECEIVED A REPORT THAT A HEATER-COOLER SYSTEM 3T WAS LEAKING WATER DURING SETUP. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1203476 HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS DWC LIVANOVA DEUTSCHLAND 16-02-80

Patients

Seq Age Sex Outcome Treatment
1