FDA Adverse Event
Malfunction
Summary report: N
HEATER-COOLER SYSTEM 3T
MDR report key: 10738904
·
Received October 27, 2020
Report
- Report Number
- 9611109-2020-00598
- Event Type
- Malfunction
- Date Received
- October 27, 2020
- Date of Event
- September 28, 2020
- Report Date
- October 26, 2020
- Manufacturer
- LIVANOVA DEUTSCHLAND
- Product Code
- DWC
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THERE WAS NO PATIENT INVOLVEMENT. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA, AND IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K191402). LIVANOVA (B)(6) MANUFACTURES THE HEATER-COOLER SYSTEM 3T . THE INCIDENT OCCURRED IN (B)(6). THE ROOT CAUSE OF THE REPORTED LEAK WAS CAUSED BY AN ASSEMBLY ERROR. THE YELLOW PLUG ON THE VALVE BLOCK OF THE CARDIOPLEGIA BRIDGE WAS MISSING.
Description of Event or Problem · 1
LIVANOVA (B)(6) RECEIVED A REPORT THAT A HEATER-COOLER SYSTEM 3T WAS LEAKING WATER DURING SETUP. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1203476 | HEATER-COOLER SYSTEM 3T | CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS | DWC | LIVANOVA DEUTSCHLAND | 16-02-80 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|