ARCHITECT I2000SR
Report
- Report Number
- 3016438761-2020-00262
- Event Type
- Malfunction
- Date Received
- October 26, 2020
- Date of Event
- October 6, 2020
- Report Date
- October 26, 2020
- Manufacturer
- ABBOTT LABORATORIES
- Product Code
- JJE
- UDI-DI
- 00380740006235
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
PATIENT IDENTIFIER = SID= (B)(6) AND NO SID FOR SECOND PATIENT. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE THE CUSTOMER REPORTED A FALSELY ELEVATED TROPONIN RESULT ON THE ARCHITECT I2000SR; SERIAL NUMBER (B)(4). THROUGH TROUBLESHOOTING THE CUSTOMER REPLACED THE VALVE, MANIFOLD KIT (ROHS) (LIST NUMBER 7-77612-03), AND ACCUMULATOR TUBE ASSEMBLY, (ROHS) PN 7-95548-01 THEN VERIFIED THE SYSTEM PERFORMED WITHIN SPECIFICATION WITHOUT ANY ISSUE. THERE HAVE BEEN NO SUBSEQUENT CONTACTS FROM THE CUSTOMER REGARDING ELEVATED TROPONIN RESULTS SINCE THE PARTS WERE REPLACED. A REVIEW OF THE I2000SR TRACKING AND TRENDING DATA FOUND NO TRENDS ASSOCIATED WITH THE CUSTOMER ISSUE DESCRIBED IN THIS COMPLAINT. A REVIEW FOR SIMILAR COMPLAINTS IDENTIFIED NO TRENDS FOR THE MANIFOLD KIT, AND ACCUMULATOR TUBE ASSEMBLY. A REVIEW OF THE MANUFACTURING DOCUMENTATION DID NOT IDENTIFY ANY ISSUES ASSOCIATED WITH THE LIKELY CAUSE PART, OR THE COMPLAINT ISSUE. LABELING WAS REVIEWED AND FOUND TO BE ADEQUATE. BASED ON THE INVESTIGATION NO PRODUCT DEFICIENCY WAS IDENTIFIED FOR THE ARCHITECT I2000SR; SERIAL NUMBER (B)(4), OR THE MANIFOLD KIT (ROHS) (LIST NUMBER 7-77612-03) AND ACCUMULATOR TUBE ASSEMBLY, (ROHS) PN 7-95548-01.
THE CUSTOMER OBSERVED FALSELY ELEVATED ARCHITECT STAT TROPONIN-I RESULTS ON MULTIPLE PATIENTS. THE RESULTS PROVIDED WERE: ON (B)(6) 2020, SID: (B)(6); INITIAL TROPONIN=0.75 NG/ML/ REPEATED ON (B)(6) 2020=NORMAL VALUE (NO ACTUAL RESULTS PROVIDED). SECOND PATIENT: 0.083 NG/ML/ (B)(6) 2020 REPEATED ON SAME INSTRUMENT =0.686 NG/ML/REPEATED ON DIFFERENT INSTRUMENT=0.089 NG/ML. NORMAL RANGE FOR TROPONIN= 0.0 TO 0.05 NG/ML, CRITICAL VALUE= >0.5 NG/ML. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1199908 | ARCHITECT I2000SR | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JJE | ABBOTT LABORATORIES | 3M74-02 | 00380740006235 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |