TRUE METRIX
Report
- Report Number
- 1000113657-2020-00781
- Event Type
- Malfunction
- Date Received
- October 26, 2020
- Date of Event
- September 29, 2020
- Report Date
- December 2, 2020
- Manufacturer
- TRIVIDIA HEALTH INC
- Product Code
- NBW
- UDI-DI
- 00021292007836
- PMA / PMN Number
- K140100
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
SECTIONS WITH ADDITIONAL INFORMATION AS OF 02-DEC-2020: H6: UPDATED FDA'S METHOD, RESULT, AND CONCLUSION CODES. H10: METER WAS RETURNED FOR EVALUATION; DEFECT FOUND ON REV 4. RETURNED METER - E-0 WITH LAB CONTROL. TEST STRIPS WERE NOT RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE: RC-079: THE CONTROL SOLUTION PEAK IS BEING DETECTED BELOW THE LOWER LIMIT OF 2.0 FOR MR2 V2.53 METER.
INTERNAL REPORT REFERENCE NUMBER: (B)(4). PRODUCTS WERE RETURNED AND PENDING QC EVALUATION. NOTE: MANUFACTURER CONTACTED CUSTOMER IN A FOLLOW-UP CALL TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - UNABLE TO ESTABLISH CONTACT WITH CUSTOMER AT THIS TIME. PRODUCT NOTIFICATION LETTER SENT TO CONTACT CUSTOMER CARE.
CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS OBTAINED OF 275 AND 398 MG/DL. THE CUSTOMER¿S EXPECTED BLOOD GLUCOSE TEST RESULT RANGES ARE: AM FASTING 150-200 MG/DL. BEFORE LUNCH FASTING LESS 225 MG/DL. BEFORE DINNER FASTING 180-211 MG/DL. THE CUSTOMER WAS NOT USING THE PROPER TESTING TECHNIQUE. THE CUSTOMER FELT WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. THE CUSTOMER DECLINED TO PERFORM A BACK TO BACK BLOOD TEST DURING THE CALL. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE LIVING AREA. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 09/19/2021 AND TEST STRIPS WERE OPENED ONE WEEK AGO. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: (B)(6).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1198112 | TRUE METRIX | SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER | NBW | TRIVIDIA HEALTH INC | STRIP, TMX CVS 100CT12/CASE MG/DL | MX4175S | 00021292007836 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |