FDA Adverse Event Injury Summary report: N

RIATA ACTIVE FIXATION LEAD

MDR report key: 1073605 · Received July 11, 2008

Report

Report Number
2017865-2008-02489
Event Type
Injury
Date Received
July 11, 2008
Date of Event
May 7, 2008
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
LWS
PMA / PMN Number
PMA00001
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE ANALYSIS OF THE LEAD DID NOT REVEAL ANY DEVICE CONDITION THAT WOULD HAVE CONTRIBUTED TO THE CAPTURE ANOMALY. VISUAL INSPECTION REVEALED AN INSULATION CUT AT 15.65 CM FROM THE CONNECTOR END, CONSISTENT WITH THAT OCCURRING AT THE TIME OF SURGICAL PROCEDURE. FUNCTIONAL TESTING REVEALED NORMAL LEAD CHARACTERISTICS. THE ELECTRICAL CHARACTERISTICS OF THE LEAD WERE FOUND TO BE NORMAL.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD WAS EXPLANTED DUE TO INTERMITTENT CAPTURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RIATA ACTIVE FIXATION LEAD DEFIBRILLATION LEAD LWS ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1580/60 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention