FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION LEAD

MDR report key: 1073592 · Received July 11, 2008

Report

Report Number
2017865-2008-02476
Event Type
Injury
Date Received
July 11, 2008
Date of Event
May 16, 2008
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
LWS
PMA / PMN Number
PMA00001
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

TESTING FOUND THAT THE LEAD STIFFNESS WAS WITHIN SPECIFICATION. THE LEAD EXHIBITED NORMAL ELECTRICAL CHARACTERISTICS. DRIED BODY FLUID/TISSUE INSIDE THE HEADER AND IN THE DISTAL INSULATION PREVENTED HELIX EXTENSION AND DIFFICULTY WITH STYLET INSERTION. HELIX FUNCTION WAS NORMAL AFTER CLEANING.

Description of Event or Problem · 1

THE LEAD WAS EXPLANTED DUE TO A LEAD PERFORATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DURATA STS OPTIM ACTIVE FIXATION LEAD DEFIBRILLATION LEAD LWS ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7120/65 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention