FDA Adverse Event
Injury
Summary report: N
TENDRIL SDX
MDR report key: 1073500
·
Received July 11, 2008
Report
- Report Number
- 2017865-2008-01995
- Event Type
- Injury
- Date Received
- July 11, 2008
- Date of Event
- April 2, 2008
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- DTB
- PMA / PMN Number
- PMA00001
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
FINAL ANALYSIS FOUND AN ABRASION ON THE PROXIMAL INSULATION AT 6.5 CM FROM THE CONNECTOR PIN, WHICH WOULD CAUSE THE REPORTED CAPTURE AND SENSING ANOMALIES.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE LEAD EXHIBITED CAPTURE AND SENSING ANOMALIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TENDRIL SDX | PERMANENT PACEMAKER ELECTRODE | DTB | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 1488T/52 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |