RIATA ACTIVE FIXATION
Report
- Report Number
- 2017865-2008-02396
- Event Type
- Injury
- Date Received
- July 11, 2008
- Date of Event
- April 18, 2008
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED THE REPORTED FIELD EVENT WAS NOT CONFIRMED. NORMAL LEAD COILS/CABLES CONTINUITIES AND LEAD INSULATION RESISTANCE WERE MEASURED ON THE RETURNED PORTION OF THE LEAD. ANALYSIS NOTED DAMAGE AT 8 - 10 CM FROM THE LEAD TIP, INDICATIVE OF LEAD ABRASION. THE APPEARANCE OF THE ABRADED SILICONE INSULATION AROUND THE EXPOSED SENSING CABLES INDICATED THAT THAT THE LEAD MOST LIKEY WAS IN A BENT POSITION, LEADING TO AN ABRASION. THE DAMAGES OBSERVED COULD HAVE CAUSED HIGH CAPTURE THRESHOLD.
AFTER SEVERAL FOLLOW-UPS WITH HIGH THRESHOLDS, PHYSICIAN DECIDED TO EXPLANT THE LEAD. LEAD DAMAGE WAS SUSPECTED. THERE WERE NO CONSEQUENCES, AND THE PATIENT WAS REPORTED TO BE IN GOOD CONDITION AFTER EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RIATA ACTIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 1580/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |