FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER PST GEL AND LITHIUM HEPARINN (LH) BLOOD COLLECTION TUBES

MDR report key: 10734488 · Received October 26, 2020

Report

Report Number
1917413-2020-01010
Event Type
Malfunction
Date Received
October 26, 2020
Date of Event
October 2, 2020
Report Date
March 18, 2021
Manufacturer
BECTON, DICKINSON & CO. (BROKEN BOW)
Product Code
JKA
UDI-DI
50382903679602
PMA / PMN Number
K945952
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR EVALUATION. THEREFORE, 10 RETENTION SAMPLES OF THE INCIDENT LOT WERE SELECTED FROM BD INVENTORY FOR EVALUATION AND UPON COMPLETION, NO ISSUES RELATING TO ERRONEOUS RESULTS WERE OBSERVED. NO DIFFICULTIES WERE ENCOUNTERED DURING BLOOD COLLECTION AND ALL TUBES APPEARED TO EXHIBIT PROPER FILL. BD WAS UNABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE (ERRONEOUS RESULTS) BECAUSE THE DEFECT WAS NOT EVIDENT IN THE TESTING OF THE CUSTOMER LOT SAMPLES. REPLICATES OF BOTH RETAIN AND CONTROL SAMPLES TESTED WERE ACCEPTABLE IN TERMS OF BOTH PRECISION AND ACCURACY. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. THIS COMPLAINT IS NOT CONFIRMED WITH RESPECT TO ERRONEOUS RESULTS. H3 OTHER TEXT : SEE H10.

Description of Event or Problem · 0

IT WAS REPORTED DURING USE THE BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARINN (LH) BLOOD COLLECTION TUBE EXPERIENCED ERRONEOUS RESULTS. THIS EVENT OCCURRED 2 TIMES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE CUSTOMER COMPLAINS "OF ELEVATED POTASSIUM LEVELS DESPITE INDICATING THEY'VE FOLLOWED ALL CORRECT PRE-ANALYTICAL STEPS¿.

Additional Manufacturer Narrative · 1

THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 0100292. MEDICAL DEVICE EXPIRATION DATE: 2021-04-30. DEVICE MANUFACTURE DATE:2020-04-09. MEDICAL DEVICE LOT #: 0133391. MEDICAL DEVICE EXPIRATION DATE: 2021-05-31. DEVICE MANUFACTURE DATE: 2020-05-12. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED DURING USE THE BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARINN (LH) BLOOD COLLECTION TUBE EXPERIENCED ERRONEOUS RESULTS. THIS EVENT OCCURRED 2 TIMES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE CUSTOMER COMPLAINS "OF ELEVATED POTASSIUM LEVELS DESPITE INDICATING THEY'VE FOLLOWED ALL CORRECT PRE-ANALYTICAL STEPS¿.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1203186 BD VACUTAINER PST GEL AND LITHIUM HEPARINN (LH) BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO. (BROKEN BOW) 367960 SEE H.10 50382903679602

Patients

Seq Age Sex Outcome Treatment
1 Other