FDA Adverse Event Injury Summary report: N

BD KIESTRA INOQULA+

MDR report key: 10733684 · Received October 26, 2020

Report

Report Number
1119779-2020-00774
Event Type
Injury
Date Received
October 26, 2020
Date of Event
September 30, 2020
Report Date
February 1, 2021
Manufacturer
BECTON, DICKINSON & CO. (SPARKS)
Product Code
JTC
UDI-DI
00382904472023
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: IT WAS REPORTED THAT THE TABLE TOP HOOD ABOVE THE SATO PRINTER OF THE INOQULA (SERIAL NUMBER (B)(6)) WOULD NOT STAY UP. ONE LAB TECHNICIAN WAS INJURED DUE TO THE COVER SLAMMING DOWN ON HER FINGERS. A FIELD SERVICE ENGINEER (FSE) WENT ONSITE AND FOUND THAT THE GAS SPRINGS ON THE HOOD WERE WORN OUT. THIS CONFIRMED THE COMPLAINT. THE FSE REPLACED THE GAS SPRINGS AND SPRING CLIPS. THIS SOLVED THE ISSUE AND THE SYSTEM WAS FUNCTIONING CORRECTLY AGAIN. CAPA (B)(4) WAS OPENED ON 30 APRIL 2019 FOR A SIMILAR ISSUE WITH A FALLING HOOD DUE TO WORN GAS SPRINGS. CORRECTION AT CUSTOMER LOCATION HAS TAKEN PLACE, BY REPLACING THE GAS SPRINGS. FURTHER CORRECTIVE ACTIONS WILL FOLLOW FROM THE CAPA.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE CHANGING THE BARCODE LABELS ON THE BD KIESTRA¿ INOQULA+¿, THE PRINTER COVER SLAMMED DOWN ON THE TECHNICIAN¿S THREE (3) MIDDLE FINGERS. THE TECHNICIAN REPORTED TO NURSE/OCCUPATIONAL FOR EVALUATION. WHEREIN IT WAS REPORTED THERE WAS NO EVIDENCE OF BROKEN BONES, BUT BRUISING WAS NOTED. NO PAIN MEDICATION WAS ADMINISTERED, NO ADDITIONAL PATIENT INFORMATION WAS REPORTED.

Additional Manufacturer Narrative · 1

MEDICAL DEVICE EXPIRATION DATE: NA. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT WHILE CHANGING THE BARCODE LABELS ON THE BD KIESTRA¿ INOQULA+¿, THE PRINTER COVER SLAMMED DOWN ON THE TECHNICIAN¿S THREE (3) MIDDLE FINGERS. THE TECHNICIAN REPORTED TO NURSE/OCCUPATIONAL FOR EVALUATION. WHEREIN IT WAS REPORTED THERE WAS NO EVIDENCE OF BROKEN BONES, BUT BRUISING WAS NOTED. NO PAIN MEDICATION WAS ADMINISTERED, NO ADDITIONAL PATIENT INFORMATION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1200528 BD KIESTRA INOQULA+ MICROBIAL SPECIMEN INOCULATION JTC BECTON, DICKINSON & CO. (SPARKS) 447202 NA 00382904472023

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention