FDA Adverse Event
Injury
Summary report: N
TENDRIL SDX
MDR report key: 1073343
·
Received July 11, 2008
Report
- Report Number
- 2017865-2008-02137
- Event Type
- Injury
- Date Received
- July 11, 2008
- Date of Event
- March 25, 2008
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- DTB
- PMA / PMN Number
- P960013
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
NO MEDWATCH FORM WAS RECEIVED. COMPETITOR.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE LEAD EXHIBITED HIGH THRESHOLDS, CAUSING THE PATIENT TO GO ASYSTOLIC.
Description of Event or Problem · 1
IT WAS REPORTED THAT LIA (LEAD INTEGRITY ALERT) TRIGGERED DUE TO OVERSENSING. T-WAVE OVERSENSING REPORTED AND R-WAVE SENSING OF 3.4 MV. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TENDRIL SDX | PERMANENT PACEMAKER ELECTRODE | DTB | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 1688TC/58 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 90 YR | Required Intervention |