FDA Adverse Event Malfunction Summary report: N

TENDRIL SDX

MDR report key: 1073274 · Received July 11, 2008

Report

Report Number
2017865-2008-02073
Event Type
Malfunction
Date Received
July 11, 2008
Date of Event
April 30, 2008
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
DTB
PMA / PMN Number
PMA00001
Removal / Correction Number
NA
Report Source
Manufacturer report
Reporter Location
KY
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT HE WAS DIZZY OVER THE PAST TWO WEEKS. ATRIAL LEAD IMPEDANCE WAS 105 OHMS. THE P-WAVES VARIED FROM 3.0 MV TO 0.3 MV WITH SUCCESSIVE TESTS. IN THE ATRIUM THERE WAS NO CAPTURE AT 7.5 V, 1.5 MS. AN X-RAY SHOWED CLAVICULAR CRUSH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TENDRIL SDX PERMANENT PACEMAKER ELECTRODE DTB ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1688TC/52 NA

Patients

Seq Age Sex Outcome Treatment
1