TENDRIL SDX
MDR report key: 1073246
·
Received July 11, 2008
Report
- Report Number
- 2017865-2008-01960
- Event Type
- Injury
- Date Received
- July 11, 2008
- Date of Event
- April 4, 2008
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- DTB
- PMA / PMN Number
- PMA00001
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE LEAD EXHIBITED LOSS OF ATRIAL CAP- TURE WITH SUBOPTIMAL ATRIAL SENSING. P WAVES DECREASED FROM GREATER THAN 2 MV AT THE LAST INTERROGATION TO 0.3 MV. THE BIPOLAR ATRIAL LEAD IMPEDANCE HAD DECREASED FROM 351 OHMS AT THE LAST CHECK, TO 319 OHMS. THE LEAD WAS CAPPED AND RE- PLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TENDRIL SDX | PERMANENT PACEMAKER ELECTRODE | DTB | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 1688TC/46 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |