FDA Adverse Event Injury Summary report: N

TENDRIL SDX

MDR report key: 1073243 · Received July 11, 2008

Report

Report Number
2017865-2008-02047
Event Type
Injury
Date Received
July 11, 2008
Date of Event
February 7, 2008
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
DTB
PMA / PMN Number
PMA00001
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

FINAL ANALYSIS FOUND THAT PRESSURE TO THE LEAD DUE TO RIB CLAVICULAR CRUSH DAMAGED THE DISTAL INSULATION AND FRACTURED THE PROXIMAL COIL, CAUSING A SHORT BETWEEN THE COILS, AND THE REPORTED LOW LEAD IMPEDANCE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD EXHIBITED LESS THAN 200 OHMS IMPEDANCE, AND AN X-RAY REVEALED THAT THE HELIX WAS BENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TENDRIL SDX PERMANENT PACEMAKER ELECTRODE DTB ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1688T/52 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention