FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION LEAD

MDR report key: 1073088 · Received July 11, 2008

Report

Report Number
2017865-2008-02334
Event Type
Injury
Date Received
July 11, 2008
Date of Event
April 4, 2008
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
LWS
PMA / PMN Number
PMA00001
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE ANALYSIS OF THE LEAD DID NOT REVEAL ANY DEVICE CONDITION THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. A LEAD TIP STIFFNESS TEST WAS PERFORMED AND WAS FOUND TO BE WITHIN SPECIFICATION. THE ELECTRICAL CHARACTERISTICS OF THE LEAD WERE FOUND TO BE NORMAL.

Description of Event or Problem · 1

EXIT BLOCK AND INTRINSIC LOW R-WAVE WERE OBSERVED. PATIENT REPORTED NOT FEELING WELL. X-RAY REVEALED A SLIGHT MOVEMENT OF THE LEAD. AS SUCH, THE LEAD WAS EXPLANTED. PATIENT WAS REPORTED TO BE WELL AFTER EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DURATA STS OPTIM ACTIVE FIXATION LEAD DEFIBRILLATION LEAD LWS ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7120/65 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention