DURATA STS OPTIM ACTIVE FIXATION LEAD
MDR report key: 1073088
·
Received July 11, 2008
Report
- Report Number
- 2017865-2008-02334
- Event Type
- Injury
- Date Received
- July 11, 2008
- Date of Event
- April 4, 2008
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- LWS
- PMA / PMN Number
- PMA00001
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE ANALYSIS OF THE LEAD DID NOT REVEAL ANY DEVICE CONDITION THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. A LEAD TIP STIFFNESS TEST WAS PERFORMED AND WAS FOUND TO BE WITHIN SPECIFICATION. THE ELECTRICAL CHARACTERISTICS OF THE LEAD WERE FOUND TO BE NORMAL.
Description of Event or Problem · 1
EXIT BLOCK AND INTRINSIC LOW R-WAVE WERE OBSERVED. PATIENT REPORTED NOT FEELING WELL. X-RAY REVEALED A SLIGHT MOVEMENT OF THE LEAD. AS SUCH, THE LEAD WAS EXPLANTED. PATIENT WAS REPORTED TO BE WELL AFTER EVENT.