SPRINT FIDELIS
Report
- Report Number
- 2649622-2008-03343
- Event Type
- Injury
- Date Received
- July 10, 2008
- Date of Event
- April 4, 2008
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- ASKU
- Removal / Correction Number
- Z-0067-007
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: PROXIMAL CONDUCTOR FRACTURED; FULL LEAD ANALYZED H3 OTHER TEXT: IT WAS REPORTED THE LEAD HAD HIGH IMPEDANCES OVER SEVERAL DAYS AND THE PATIENT FELT SYNCOPAL AND DIZZY. THE LEAD WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT REPORTED SHE WAS DOING WELL. ACTUAL DEVICE INVOLVED IN INCIDENT WAS EVALUATED ELECTRICAL TESTS PERFORMED, MECHANICAL TESTS PERFORMED, VISUAL EXAMINATION COMPONENT/SUBASSEMBLY FAILURE LEAD CONDUCTOR DEVICE WAS OUT OF SPECIFICATION IN A MANNER THAT RELATES TO EVENT IMPEDANCE, HIGH.
IT WAS REPORTED THE LEAD HAD HIGH IMPEDANCES OVER SEVERAL DAYS AND THE PATIENT FELT SYNCOPAL AND DIZZY. THE LEAD WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT REPORTED SHE WAS DOING WELL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT FIDELIS | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6949 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 27 YR | Required Intervention | 4076 IMPLANTABLE PACING LEAD| D154ATG IMPLANTABLE PACEMAKER/CARDIO/DEFIB |