LIGACLIP ENDOSCOPIC MULTIPLE CLIP APPLIER
Report
- Report Number
- 1527736-1997-01732
- Event Type
- Malfunction
- Date Received
- July 21, 1997
- Report Date
- May 29, 1997
- Manufacturer
- ETHICON ENDO-SURGERY, INC.
- Product Code
- GDO
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO
- Reporter Occupation
- PHYSICIAN
Narratives
D5,6;H3,4,6;G4: ADDED ADDITIONAL INFO. D6: DEVICE RETURNED WITH NO LOT ID. VISUAL INSPECTIONS & RESULTS: CLIP IN JAWS, AEFGH-NO BCD-YES; DAMAGED JAWS, DAMAGED FEED BAR, DAMAGED FLOOR, DAMAGED HANDLE SHROUDS, DAMAGED TIP SHROUDS, DAMAGED TRIGGER, DAMAGED TUBE, AND DAMAGED WELD SEAMS, A-H NO; DAMAGED CUTTER, NA; AND DAMAGED OTHER, AE-BROKEN TOP AND. FUNCTIONAL TESTS & RESULTS: ANTIBACKUP FUNCTIONAL, FIRING: FEED AND FORM, JAWS: HOLD CLIP, AND LOCKOUT FUNCTIONAL, ACDFGH-YES BE-NA; AND NUMBER OF CLIPS FED AND FORMED, A-11 C-6 D-3 FGH-21. ANALYSIS CONCLUSION: BASED ON THE INFO RECEIVED AND THE VISUAL AND FUNCTIONAL INSPECTION, NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THERE WERE EIGHT INSTRUMENTS RETURNED, THREE OF WHICH WERE IN THE STERILE BLISTER. TWO OF THE INSTRUMENTS WERE RETURNED EMPTY AND LOCKED OUT, THEREFORE FURTHER FUNCTIONAL TESTING COULD NOT BE PERFORMED ON THE EMPTY INSTRUMENTS. THE REMAINING SIX INSTRUMENTS WERE FIRED AND ALL FIRED AND FORMED THE CLIPS AS DESIGNED WITH NO DIFFICULTIES NOTED. IT COULD NOT BE ASCERTAINED AS TO WHY THE CLIPS REPORTEDLY MALFORMED. EACH INSTRUMENT IS EVALUATED DURING THE ASSEMBLY PROCESS TO ENSURE THE CLIPS FEED AND FORM PROPERLY.
IT WAS REPORTED THE DEVICES WERE USED DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE. IT WAS REPORTED BY THE AFFILIATE THAT THE CLIPS WERE MALFORMED. THIS CAUSED THE CLIPS TO FALL FROM THE JAWS OF THE DEVICES. THERE WAS NO PT CONSEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGACLIP ENDOSCOPIC MULTIPLE CLIP APPLIER | ENDOSCOPIC CLIP APPLIER | GDO | ETHICON ENDO-SURGERY, INC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |