FDA Adverse Event Injury Summary report: N

CONCERTO CRT-D DR

MDR report key: 1072680 · Received July 10, 2008

Report

Report Number
2647346-2008-00349
Event Type
Injury
Date Received
July 10, 2008
Date of Event
May 21, 2008
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
NIK
PMA / PMN Number
ASKU
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: ACTUAL LONGEVITY IS < 80% OF 99.9% LONGEVITY LIMIT; THE DEVICE WAS FULLY FUNCTIONAL, WITH NO HIGH CURRENT DRAIN OR EVIDENCE OF BATTERY PROBLEMS. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONCERTO CRT-D DR IMPLANTABLE PACEMAKER/CARDIO/DEFIB NIK MEDTRONIC MED REL, INC. C154DWK ASKU

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention 6947 IMPLANTABLE TACHY LEAD| 4076 IMPLANTABLE PACING LEAD