FDA Adverse Event Malfunction Summary report: N

HEATER-COOLER SYSTEM 3T

MDR report key: 10725957 · Received October 23, 2020

Report

Report Number
9611109-2020-00595
Event Type
Malfunction
Date Received
October 23, 2020
Date of Event
September 24, 2020
Report Date
October 22, 2020
Manufacturer
LIVANOVA DEUTSCHLAND
Product Code
DWC
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THERE WAS NO PATIENT INVOLVEMENT. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA AND IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K191402). LIVANOVA (B)(4) MANUFACTURES THE HEATER-COOLER SYSTEM 3T . THE INCIDENT OCCURRED IN (B)(6). A LIVANOVA FIELD SERVICE REPRESENTATIVE WAS DISPATCHED TO THE FACILITY TO INVESTIGATE AND CONFIRMED THE REPORTED ISSUE. THE FAULT WAS TRACED BACK TO A DEFECTIVE PATIENT PUMP 2 CIRCUIT BOARD. THE PUMP WAS REPLACED AND THE ISSUE SOLVED. SUBSEQUENT FUNCTIONAL VERIFICATION TESTING WAS COMPLETED WITHOUT FURTHER ISSUES AND THE UNIT WAS RETURNED TO SERVICE.

Description of Event or Problem · 1

LIVANOVA (B)(4) RECEIVED A REPORT THAT A HEATER-COOLER SYSTEM 3T DISPLAYED AN ERROR CODE ASSOCIATED TO ONE PATIENT PUMP DURING SERVICE. THERE WAS NO REPORT OF PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1192662 HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS DWC LIVANOVA DEUTSCHLAND 16-02-80

Patients

Seq Age Sex Outcome Treatment
1