8015 ALARIS SYSTEM PC UNIT
Report
- Report Number
- 2016493-2020-25625
- Event Type
- Malfunction
- Date Received
- October 23, 2020
- Report Date
- March 23, 2018
- Manufacturer
- CAREFUSION SD
- Product Code
- FRN
- PMA / PMN Number
- K091308
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
THIS REPORTED EVENT AND SUBSEQUENT REPAIRS WERE INVESTIGATED THROUGH THE SERVICE REPAIR PROCESS. FAILURE DATA AND PARTS-USED INFORMATION WERE REVIEWED FOR THE SAP AND TRACKWISE FILES AND FOUND RELEVANT TO THE SERVICE REPAIR. A REVIEW OF THE DEVICE SERVICE HISTORY RECORD WAS PERFORMED FROM THE DATE OF MANUFACTURE TO THE DATE CORRESPONDING TO THIS SERVICE NOTIFICATION NUMBER. THE DEVICE WAS NOT PREVIOUSLY RETURNED FOR SERVICE. THE DATABASE SHOWED NO QUALITY NOTIFICATIONS WERE OPENED FOR THE DEVICE. A REVIEW OF THE DEVICE HISTORY RECORD IN SAP FOR SN (B)(4) WAS PERFORMED WHICH CONFIRMED THAT THIS DEVICE WAS NOT INVOLVED IN A PRODUCTION FAILURE, AND PRODUCT WAS RETURNED FOR SERVICING WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. A REVIEW OF THE COMPLAINT HISTORY RECORD IN THE TRACKWISE WAS PERFORMED FOR THE SN (B)(4) WHICH CONFIRMED NO SIMILAR COMPLAINTS WITH THE SAME OR RELATED FAILURE MODE. CAPA REFERENCE: (B)(4). THE CUSTOMER STATED THAT THERE WAS NO PATIENT INVOLVEMENT.
(B)(4). INVESTIGATION SUMMARY: REPORTED 'TAMPER SWITCH INOP' DURING THE DEVICE CHECK IN WAS CONFIRMED. AS RECEIVED, THE TAMPER SWITCH IS CONFIRMED FAILING DURING THE KEYPAD TESTING. THE SIO BOARD ASSEMBLY WAS REMOVED FOR VISUAL INSPECTION OF THE TAMPER SWITCH ASSEMBLY. NO ANOMALIES' OBSERVED, THE TAMPER SWITCH HARNESS WERE PROPERLY SOLDERED AND THE SWITCH IS BOUNCING PROPERLY. FAILURE INVESTIGATION WAS ABLE TO REPRODUCE REPORT 'TAMPER SWITCH INOPERATE' TO THE FASTENING OF THE SIO BOARD. CONCLUSION: FAILURE INVESTIGATION WAS ABLE TO REPRODUCE REPORT 'TAMPER SWITCH INOPERATE' TO THE FASTENING OF THE SIO BOARD. FOR BETTER RESULT, IT RECOMMENDING TIGHTENED A SCREW IN CROSS PATTERN. REWORK: RE-INSTALLED SIO BOARD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1192958 | 8015 ALARIS SYSTEM PC UNIT | PUMP, INFUSION | FRN | CAREFUSION SD | 8015 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |