FDA Adverse Event Malfunction Summary report: N

8015 ALARIS SYSTEM PC UNIT

MDR report key: 10725365 · Received October 23, 2020

Report

Report Number
2016493-2020-25625
Event Type
Malfunction
Date Received
October 23, 2020
Report Date
March 23, 2018
Manufacturer
CAREFUSION SD
Product Code
FRN
PMA / PMN Number
K091308
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

THIS REPORTED EVENT AND SUBSEQUENT REPAIRS WERE INVESTIGATED THROUGH THE SERVICE REPAIR PROCESS. FAILURE DATA AND PARTS-USED INFORMATION WERE REVIEWED FOR THE SAP AND TRACKWISE FILES AND FOUND RELEVANT TO THE SERVICE REPAIR. A REVIEW OF THE DEVICE SERVICE HISTORY RECORD WAS PERFORMED FROM THE DATE OF MANUFACTURE TO THE DATE CORRESPONDING TO THIS SERVICE NOTIFICATION NUMBER. THE DEVICE WAS NOT PREVIOUSLY RETURNED FOR SERVICE. THE DATABASE SHOWED NO QUALITY NOTIFICATIONS WERE OPENED FOR THE DEVICE. A REVIEW OF THE DEVICE HISTORY RECORD IN SAP FOR SN (B)(4) WAS PERFORMED WHICH CONFIRMED THAT THIS DEVICE WAS NOT INVOLVED IN A PRODUCTION FAILURE, AND PRODUCT WAS RETURNED FOR SERVICING WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. A REVIEW OF THE COMPLAINT HISTORY RECORD IN THE TRACKWISE WAS PERFORMED FOR THE SN (B)(4) WHICH CONFIRMED NO SIMILAR COMPLAINTS WITH THE SAME OR RELATED FAILURE MODE. CAPA REFERENCE: (B)(4). THE CUSTOMER STATED THAT THERE WAS NO PATIENT INVOLVEMENT.

Description of Event or Problem · 1

(B)(4). INVESTIGATION SUMMARY: REPORTED 'TAMPER SWITCH INOP' DURING THE DEVICE CHECK IN WAS CONFIRMED. AS RECEIVED, THE TAMPER SWITCH IS CONFIRMED FAILING DURING THE KEYPAD TESTING. THE SIO BOARD ASSEMBLY WAS REMOVED FOR VISUAL INSPECTION OF THE TAMPER SWITCH ASSEMBLY. NO ANOMALIES' OBSERVED, THE TAMPER SWITCH HARNESS WERE PROPERLY SOLDERED AND THE SWITCH IS BOUNCING PROPERLY. FAILURE INVESTIGATION WAS ABLE TO REPRODUCE REPORT 'TAMPER SWITCH INOPERATE' TO THE FASTENING OF THE SIO BOARD. CONCLUSION: FAILURE INVESTIGATION WAS ABLE TO REPRODUCE REPORT 'TAMPER SWITCH INOPERATE' TO THE FASTENING OF THE SIO BOARD. FOR BETTER RESULT, IT RECOMMENDING TIGHTENED A SCREW IN CROSS PATTERN. REWORK: RE-INSTALLED SIO BOARD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1192958 8015 ALARIS SYSTEM PC UNIT PUMP, INFUSION FRN CAREFUSION SD 8015

Patients

Seq Age Sex Outcome Treatment
1