Report Number
2016493-2020-25857
Event Type
Malfunction
Date Received
October 22, 2020
Report Date
May 8, 2018
Manufacturer
CAREFUSION SD
Product Code
FRN
UDI-DI
10885403801518
PMA / PMN Number
K091308
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

THIS REPORTED EVENT AND SUBSEQUENT REPAIRS WERE INVESTIGATED THROUGH THE TSC TROUBLESHOOTING PROCESS. A REVIEW OF THE DEVICE SERVICE HISTORY RECORD WAS PERFORMED FROM THE DATE OF MANUFACTURE TO THE DATE CORRESPONDING TO THIS SERVICE NOTIFICATION NUMBER. THE DEVICE WAS NOT RETURNED FOR SERVICE. THE DATABASE SHOWED NO QUALITY NOTIFICATIONS WERE OPENED FOR THE DEVICE. A REVIEW OF THE DEVICE HISTORY RECORD IN SAP FOR SN (B)(4) WAS PERFORMED WHICH CONFIRMED THAT THIS DEVICE WAS NOT INVOLVED IN A PRODUCTION FAILURE, AND PRODUCT WAS NOT RETURNED FOR SERVICING WHICH NOT CORRELATES TO THE CUSTOMER REPORTED ISSUE. A REVIEW OF THE COMPLAINT HISTORY RECORD IN THE TRACKWISE WAS PERFORMED FOR THE SN (B)(4) WHICH CONFIRMED NO SIMILAR COMPLAINTS WITH THE SAME OR RELATED FAILURE MODE. CAPA REFERENCE : (B)(4). THE CUSTOMER STATED THAT THERE WAS NO PATIENT INVOLVEMENT.

Description of Event or Problem · 1

(B)(4). PLEASE REFER TO FILE #2018-016768-242819 FROM CAD. (B)(4). NPI SN (B)(4). (B)(6) HAD A PCU8015 FAILED BATTERY STATUS CHECK IN ASM SOFTWARE DURING PM. I TOLD HIM TO FOLLOW SERVICE BULLETIN 592A AND PERFORM MANUAL BATTERY CONDITIONING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1180785 8015 ALARIS SYSTEM PC UNIT PUMP, INFUSION FRN CAREFUSION SD 8015 10885403801518

Patients

Seq Age Sex Outcome Treatment
1