PEN NDL 32G 6MM 3B TW 70CT JPN
Report
- Report Number
- 9616656-2020-01053
- Event Type
- Malfunction
- Date Received
- October 22, 2020
- Date of Event
- October 2, 2020
- Report Date
- November 12, 2020
- Manufacturer
- BECTON DICKINSON AND CO.
- Product Code
- FMI
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H.6. INVESTIGATION: FOUR OPEN AND 63 SEALED 32G X 6MM PEN NEEDLE SAMPLES AND FOUR PHOTOS WERE RETURNED FROM LOT. NO. 9121516, CAT. NO. 320738. A CLOG TEST WAS CARRIED OUT ON THE FOUR OPEN SAMPLES AND TWENTY SIX SEALED SAMPLES AS PER TP700483 AND NO ISSUES WERE OBSERVED. A LOT HISTORY REVIEW WAS CARRIED OUT AND NO RELATED NON CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS PACKAGED LOT CONCLUDING ALL INSPECTIONS WERE PERFORMED AS PER THE APPLICABLE OPERATIONS AND MET QC SPECIFICATIONS. NO ISSUES WERE OBSERVED WITH THE RETURNED SAMPLES THEREFORE NO ROOT CAUSE CAN BE IDENTIFIEDSEE H.10
IT WAS REPORTED THAT 2 PEN NDL 32G 6MM 3B TW 70CT JPN WERE UNABLE/DIFFICULT TO PRIME PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER; WHEN PRIMING, INSULIN DIDN'T COME OUT AT ALL. THE INSULIN IS NOVO'S RYZODEG. THE CUSTOMER HAS ALREADY ASKED NOVO TO INVESTIGATE THE PEN. AT THE PHARMACY, THE PHARMACIST ATTEMPTED TO OPERATE THE PEN NEEDLE TOGETHER WITH THE PATIENT AND CONFIRMED THAT DRUG DIDN'T COME OUT WITH 2 PEN NEEDLES.
A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).
IT WAS REPORTED THAT 2 PEN NDL 32G 6MM 3B TW 70CT JPN WERE UNABLE/DIFFICULT TO PRIME PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER; WHEN PRIMING, INSULIN DIDN'T COME OUT AT ALL. THE INSULIN IS NOVO'S RYZODEG. THE CUSTOMER HAS ALREADY ASKED NOVO TO INVESTIGATE THE PEN. AT THE PHARMACY, THE PHARMACIST ATTEMPTED TO OPERATE THE PEN NEEDLE TOGETHER WITH THE PATIENT AND CONFIRMED THAT DRUG DIDN'T COME OUT WITH 2 PEN NEEDLES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1188078 | PEN NDL 32G 6MM 3B TW 70CT JPN | PEN NEEDLE | FMI | BECTON DICKINSON AND CO. | 9121516 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |