SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2008-03918
- Event Type
- Injury
- Date Received
- July 10, 2008
- Date of Event
- May 30, 2008
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- ASKU
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK
- Reporter Occupation
- INVALID DATA
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: BLOOD IN/ON HELIX/LOBE MECHANISM; FULL LEAD WAS RETURNED FOR ANALYSIS. THE LEAD WAS RETURNED WITH A LOT OF BLOOD IN AND ON THE HELIX AREA THAT MOST LIKELY PREVENTED THE HELIX FROM FUNCTIONING PROPERLY. THERE WERE NO ANOMALIES FOUND THAT WOULD CONTRIBUTE TO THE REPORTED ELECTRICAL ISSUES. IT WAS REPORTED THE LEAD HAD DISLODGED, WITH SENSING DIFFICULTY, FIXATION DIFFICULTY, HIGH THRESHOLDS, NO CAPTURE AND THERE WAS A PERFORATION. AT THE TIME OF LEAD REVISION ATTEMPT, THERE WAS DIFFICULTY EXTENDING/RETRACTING THE HELIX AND THE LEAD WAS REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. ACTUAL DEVICE INVOLVED IN INCIDENT WAS EVALUATED ELECTRICAL TESTS PERFORMED, MECHANICAL TESTS PERFORMED, VISUAL EXAMINATION COMPONENT/SUBASSEMBLY FAILURE LEAD CONDUCTOR OPERATIONAL CONTEXT CONTRIBUTED TO EVENT CAPTURE, FAILURE TO POSITIONING DIFFICULTIES SENSITIVITY HIGH THRESHOLD.
IT WAS REPORTED THE LEAD HAD DISLODGED, WITH SENSING DIFFICULTY, FIXATION DIFFICULTY, HIGH THRESHOLDS, NO CAPTURE AND THERE WAS A PERFORATION. AT THE TIME OF LEAD REVISION ATTEMPT, THERE WAS DIFFICULTY EXTENDING/RETRACTING THE HELIX AND THE LEAD WAS REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | 6947 IMPLANTABLE TACHY LEAD| D154VWC IMPLANTABLE PACEMAKER/CARDIO/DEFIB |