FDA Adverse Event Injury Summary report: N

SPRINT

MDR report key: 1072237 · Received July 10, 2008

Report

Report Number
2649622-2008-03573
Event Type
Injury
Date Received
July 10, 2008
Date of Event
April 15, 2008
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
ASKU
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: DISTAL CONDUCTOR FRACTURED; PARTIAL LEAD IN SEGMENTS RETURNED AND ANALYZED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6943 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Other| R