DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3004753838-2020-124231
- Event Type
- Malfunction
- Date Received
- October 22, 2020
- Date of Event
- September 23, 2020
- Report Date
- December 7, 2020
- Manufacturer
- DEXCOM, INC.
- Product Code
- QBJ
- UDI-DI
- 00386270000385
- PMA / PMN Number
- DEN170088
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
COM-1961309-20. B5: DESCRIBE EVENT OR PROBLEM - ADDITIONAL. D4: ADDITIONAL DEVICE INFORMATION - ADDITIONAL. D10: DEVICE AVAILABLE FOR EVALUATION- ADDITIONAL . H2: ADDITIONAL INFORMATION/DEVICE EVALUATION. H3: DEVICE EVALUATED BY MANUFACTURER - ADDITIONAL. H6: EVENT PROBLEM AND EVALUATION CODES - ADDITIONAL.
IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. TRANSMITTER VOLTAGE TEST WAS PERFORMED AND PASSED DUE TO (0.21 VDC). NORDIC BLUETOOTH PAIRING TEST WAS PERFORMED AND PASSED. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND SIGNAL LOSS WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS NOT CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
(B)(4).
IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1186887 | DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM | CONTINUOUS GLUCOSE MONITOR | QBJ | DEXCOM, INC. | 9445-02 | 7270539 | 00386270000385 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR |