ADAPTA DR
Report
- Report Number
- 2647346-2008-00356
- Event Type
- Malfunction
- Date Received
- July 10, 2008
- Date of Event
- May 14, 2008
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- DXY
- PMA / PMN Number
- ASKU
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: NO ANOMALIES WERE FOUND. IT WAS REPORTED THAT THE DEVICE HAD SENSING DIFFICULTY AND HIGH ATRIAL IMPEDANCE. THE DEVICE WAS EXPLANTED AND REPLACED. THERE WERE NO REPORTED PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. ACTUAL DEVICE INVOLVED IN INCIDENT WAS EVALUATED ELECTRICAL TESTS PERFORMED, MECHANICAL TESTS PERFORMED, VISUAL EXAMINATION DEVICE PERFORMED ACCORDING TO SPECIFICATIONS DEVICE EVALUATED AND ALLEGED FAILURE COULD NOT BE DUPLICATED IMPEDANCE, HIGH SENSITIVITY.
IT WAS REPORTED THAT THE DEVICE HAD SENSING DIFFICULTY AND HIGH ATRIAL IMPEDANCE. THE DEVICE WAS EXPLANTED AND REPLACED. THERE WERE NO REPORTED PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADAPTA DR | IMPLANTABLE PULSE GENERATOR | DXY | MEDTRONIC MED REL, INC. | ADDRL1 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | 4076 IMPLANTABLE PACING LEAD |