FDA Adverse Event
Injury
Summary report: N
ATTAIN OTW
MDR report key: 1072052
·
Received July 10, 2008
Report
- Report Number
- 2649622-2008-03832
- Event Type
- Injury
- Date Received
- July 10, 2008
- Date of Event
- January 10, 2008
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWP
- PMA / PMN Number
- ASKU
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY - THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. NO ANOMALIES FOUND WAS NOTED ON THE DATA DISC REVIEW.
Additional Manufacturer Narrative · 1
CORRECTED: PRODUCT EVENT SUMMARY: THE FULL LEAD IN SEGMENTS WAS RETURNED, ANALYZED, AND THE OUTER INSULATION OF THE LEAD DEVELOPED A BREACH DUE TO ENVIRONMENTAL STRESS CRACKING WHILE IN VIVO. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
Description of Event or Problem · 1
ASKU
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ATTAIN OTW | PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE | LWP | MEDTRONIC PUERTO RICO, INC. | 4193 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | Required Intervention | 4076 IMPLANTABLE PACING LEAD| 7289 IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 6949 IMPLANTABLE TACHY LEAD |