FDA Adverse Event Injury Summary report: N

INSYNC II MARQUIS

MDR report key: 1071972 · Received July 10, 2008

Report

Report Number
6000094-2008-00259
Event Type
Injury
Date Received
July 10, 2008
Date of Event
December 11, 2007
Manufacturer
MEDTRONIC S.A.
Product Code
NIK
PMA / PMN Number
ASKU
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: ACTUAL LONGEVITY IS LESS THAN 80% OF 99.9% LONGEVITY LIMIT THE DEVICE WAS FULLY FUNCTIONAL, WITH NO HIGH CURRENT DRAIN OR EVIDENCE OF BATTERY PROBLEMS. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.

Description of Event or Problem · 1

ASKU

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INSYNC II MARQUIS IMPLANTABLE PACEMAKER/CARDIO/DEFIB NIK MEDTRONIC S.A. 7289 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention 5076 IMPLANTABLE PACING LEAD| 6948 IMPLANTABLE TACHY LEAD| 4194 IMPLANTABLE PACING LEAD