FDA Adverse Event Injury Summary report: N

ONYX VR

MDR report key: 1071947 · Received July 10, 2008

Report

Report Number
6000094-2008-00289
Event Type
Injury
Date Received
July 10, 2008
Date of Event
August 21, 2007
Manufacturer
MEDTRONIC S.A.
Product Code
LWS
PMA / PMN Number
ASKU
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: ACTUAL LONGEVITY IS < 80% OF 99.9% LONGEVITY LIMIT. THE DEVICE WAS FULLY FUNCTIONAL, WITH NO HIGH CURRENT DRAIN OR EVIDENCE OF BATTERY PROBLEMS. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONYX VR IMPLANTABLE PACEMAKER/CARDIO/DEFIB LWS MEDTRONIC S.A. 7290CX ASKU

Patients

Seq Age Sex Outcome Treatment
1 Other| R 6947 IMPLANTABLE TACHY LEAD