FDA Adverse Event
Malfunction
Summary report: N
LIFEPAK 20 DEFIBRILLATOR/MONITOR
MDR report key: 1071858
·
Received July 8, 2008
Report
- Report Number
- 3015876-2008-00775
- Event Type
- Malfunction
- Date Received
- July 8, 2008
- Date of Event
- June 17, 2008
- Report Date
- June 17, 2008
- Manufacturer
- PHYSIO-CONTROL, INC.
- Product Code
- LDD
- PMA / PMN Number
- K012274
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
PHYSIO-CONTROL EVALUATED THE DEVICE AND OBSERVED THAT IT WOULD NOT OPERATE ON BATTERY POWER. PHYSIO REPLACED THE POWER SUPPLY ASSEMBLY AND THEN OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE. PHYSIO FURTHER EVALUATED THE REPLACED POWER SUPPLY ASSEMBLY, AND DETERMINED THAT ROOT CAUSE FOR THE REPORTED INCIDENT WAS AN ELECTRICALLY LEAKY FILTER, DESIGNATOR FL4, DUE TO SOLDER RESIDUE ON THE PCB.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE FAILED TO POWER ON USING A BATTERY POWER. THERE WAS NO PT USE ASSOCIATED WITH THE REPORTED EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEPAK 20 DEFIBRILLATOR/MONITOR | LDD | PHYSIO-CONTROL, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |