FDA Adverse Event Other Summary report: N

MEDCARE STAND-N-WEIGHT

MDR report key: 1071745 · Received May 6, 2008

Report

Report Number
2135150-2008-00002
Event Type
Other
Date Received
May 6, 2008
Date of Event
April 26, 2008
Report Date
May 6, 2008
Manufacturer
MEDCARE PRODUCTS, INC.
Product Code
FNG
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE SHEARING OF THE PIN IS LIKELY CAUSE BY LATERAL PRESSURE ON THE ACTUATOR. THIS COULD HAVE BEEN CAUSED BY IMPACT TO THE MOTOR, LATERAL PRESSURE OVER TIME OR SOMETHING PLACED BETWEEN THE ACTUATOR AND THE STAND MAST. THIS STAND WAS 8.5 YEARS OLD. THE ACTUATOR WAS REPLACED AND MAINTENANCE INSTRUCTED ABOUT THE IMPORTANCE OF AVOIDING LATERAL PRESSURE TO THE MOTOR AND TO TAKE THE STAND OUT OF SERVICE IF IT MAKES IRREGULAR NOISES OR SHAKES OR HAS A JERKING MOTION. SALES REP IS SCHEDULED TO MAKE AN IN SERVICE VISIT.

Description of Event or Problem · 1

DURING A TRANSFER, THE LOWER PART OF THE SCREW PIN ON THE ACTUATOR BENT AND SHEARED OFF OF THE MOTOR. THE RESIDENT WAS LOWERED TO THE GROUND WITH THE ASSISTANCE OF STAFF.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MEDCARE STAND-N-WEIGHT SIT TO STAND LIFT FNG MEDCARE PRODUCTS, INC. 400005

Patients

Seq Age Sex Outcome Treatment
1 Other