FDA Adverse Event
Other
Summary report: N
MEDCARE STAND-N-WEIGHT
MDR report key: 1071745
·
Received May 6, 2008
Report
- Report Number
- 2135150-2008-00002
- Event Type
- Other
- Date Received
- May 6, 2008
- Date of Event
- April 26, 2008
- Report Date
- May 6, 2008
- Manufacturer
- MEDCARE PRODUCTS, INC.
- Product Code
- FNG
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE SHEARING OF THE PIN IS LIKELY CAUSE BY LATERAL PRESSURE ON THE ACTUATOR. THIS COULD HAVE BEEN CAUSED BY IMPACT TO THE MOTOR, LATERAL PRESSURE OVER TIME OR SOMETHING PLACED BETWEEN THE ACTUATOR AND THE STAND MAST. THIS STAND WAS 8.5 YEARS OLD. THE ACTUATOR WAS REPLACED AND MAINTENANCE INSTRUCTED ABOUT THE IMPORTANCE OF AVOIDING LATERAL PRESSURE TO THE MOTOR AND TO TAKE THE STAND OUT OF SERVICE IF IT MAKES IRREGULAR NOISES OR SHAKES OR HAS A JERKING MOTION. SALES REP IS SCHEDULED TO MAKE AN IN SERVICE VISIT.
Description of Event or Problem · 1
DURING A TRANSFER, THE LOWER PART OF THE SCREW PIN ON THE ACTUATOR BENT AND SHEARED OFF OF THE MOTOR. THE RESIDENT WAS LOWERED TO THE GROUND WITH THE ASSISTANCE OF STAFF.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MEDCARE STAND-N-WEIGHT | SIT TO STAND LIFT | FNG | MEDCARE PRODUCTS, INC. | 400005 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |