FDA Adverse Event Malfunction Summary report: N

CANON

MDR report key: 10714503 · Received October 21, 2020

Report

Report Number
2020563-2020-00006
Event Type
Malfunction
Date Received
October 21, 2020
Date of Event
September 22, 2020
Report Date
November 17, 2023
Manufacturer
CANON MEDICAL SYSTEMS CORPORATION
Product Code
LNH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION CONFIRMED THAT THE TERMINAL BLOCK WAS SEVERELY BURNED AND THAT THE BOLTS WERE LOOSENED. IN ADDITION, IT WAS DISCOVERED THAT THE BOLT IN QUESTION HAD BEEN REMOVED DURING SERVICE AND THERE WAS AN ISSUE WITH THE SERVICE PERFORMED. THE CONTACT RESISTANCE INCREASED OVER TIME, AND HEAT GENERATION AND DISCHARGE CAUSED THE RESIN BLOCK OF THE TERMINAL TO MELT AND BURN AND FIRE SPREAD TO THE COVER PLATE. THE COVER PLATE WAS DESIGNED WITH A FLAME-RETARDANT GRADE OF V-0 BUT WAS INSTALLED WITH A LOWER FLAME-RETARDANT GRADE EQUIVALENT TO HB CAUSING THE FIRE TO SPREAD. THE BOLTS ON THE TERMINAL BLOCK IN QUESTION ARE MANAGED AS IMPORTANT FASTENING PARTS, AND NORMALLY A TORQUE WRENCH IS USED. OVER THE COURSE OF THE INVESTIGATION WE DISCOVERED THAT THE PROPER TORQUE WRENCH WAS NOT USED AT THIS SITE. THEREFORE, WE INSTRUCTED THE SERVICE TEAM TO THOROUGHLY CARRY OUT THE WORK IN (B)(6) 2021. REGARDING THE USE OF HB FLAME RETARDANT MATERIALS IN EQUIPMENT, WE DECIDED TO USE FLAME RETARDANT GRADE V-0 AND CHANGED THE PROCESS TO MANAGE THEM AS QUALITY-CRITICAL PARTS. FOR EXISTING SITES, WE REPLACED THE PLATE WITH THE CORRECT FLAME-RETARDANT MATERIAL. DURING THE REPLACEMENT WORK, WE CONFIRMED THAT THE BOLTS ON THE TERMINALS WERE TIGHTENED, THAT THERE WERE NO FOREIGN OBJECTS CAUGHT IN THEM AND THAT THE CABLES WERE PROPERLY SECURED. WE EXPLAINED THE NATURE OF THE PROBLEM TO CUSTOMERS USING THE EQUIPMENT SHIPPED BY OUR COMPANY AND COMPLETED THE REPAIR BY CHANGING THE PARTS TO SPECIFICATIONS AND INSPECTING THE DEVICES.

Description of Event or Problem · 0

CUSTOMER ALLEGED THAT MRI TECH WAS PERFORMING A HEAD SCAN AND ENTERED MRI SUITE TO INJECT PATIENT FOR THE LAST FIVE MINUTES OF THE SCAN. TECH ALLEGES THAT SHE DID NOT NOTICE ANYTHING OUT OF THE ORDINARY IN MRI SUITE AT THIS TIME. AFTER THE SCAN HAD COMPLETED AND WHILE RETRIEVING PATIENT, THE TECH ALLEGED THAT SHE SMELLED SMOKE AND UPON INVESTIGATION, SAW SMOKE NEAR THE BELLOWS. TECH STATES THAT SHE PROMPTLY REMOVED PATIENT FROM MRI SUITE TO A SAFE DISTANCE.

Additional Manufacturer Narrative · 1

DUE TO COVID-19, THE INVESTIGATION AT THE SITE WITH ALL AFFECTED PARTIES HAS BEEN DELAYED AND THE MANUFACTURER WILL SUBMIT A FOLLOW-UP REPORT UPON COMPLETION OF THE INVESTIGATION.

Description of Event or Problem · 1

CUSTOMER ALLEGED THAT MRI TECH WAS PERFORMING A HEAD SCAN AND ENTERED MRI SUITE TO INJECT PATIENT FOR THE LAST FIVE MINUTES OF THE SCAN. TECH ALLEGES THAT SHE DID NOT NOTICE ANYTHING OUT OF THE ORDINARY IN MRI SUITE AT THIS TIME. AFTER THE SCAN HAD COMPLETED AND WHILE RETRIEVING PATIENT, THE TECH ALLEGED THAT SHE SMELLED SMOKE AND UPON INVESTIGATION, SAW SMOKE NEAR THE BELLOWS. TECH STATES THAT SHE PROMPTLY REMOVED PATIENT FROM MRI SUITE TO A SAFE DISTANCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1180298 CANON MAGNETIC RESONANCE DIAGNOSTIC DEVICE LNH CANON MEDICAL SYSTEMS CORPORATION MRT-1510/U4

Patients

Seq Age Sex Outcome Treatment
1 Unknown