MARYLAND BIPOLAR FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2008-01128
- Event Type
- Malfunction
- Date Received
- July 7, 2008
- Report Date
- July 7, 2008
- Manufacturer
- INTUITIVE SURGICAL, INC.
- Product Code
- NAY
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- UNKNOWN
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER'S REPORTED COMPLAINT, ENGINEERING CONDUCTED PERFORMANCE TESTING AND FOUND THE INSTRUMENT TO PASS RECOGNITION AND ENGAGEMENT TESTS. THE INSTRUMENT MOVED INTUITIVELY WITH A FULL RANGE OF MOTION IN ALL DIRECTIONS WITH THE GRIPS OPENED AND CLOSED PROPERLY. ENGINEERING ALSO FOUND THE DISTAL END OF THE MAIN TUBE TO HAVE A 3.5" LONG SECTION WITH MATERIAL REMOVED ON ONE SIDE OF THE TUBE. THE DAMAGED AREA IS PARALLEL TO THE TUBE AXIS AND HAS A ROUGH SURFACE FINISH. BASED ON THE LOCATION AND APPEARANCE, THE DAMAGE WAS MOST LIKELY CAUSED BY A CANNULA ACCESSORY. ADDITIONAL INVESTIGATION IS UNDERWAY REGARDING THE CANNULA ACCESSORIES. A FOLLOW-UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION IS RECEIVED.
IT WAS REPORTED THAT THE MARYLAND BIPOLAR FORCEPS INSTRUMENT DID NOT WORK. NO ADDITIONAL INFORMATION WAS PROVIDED. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MARYLAND BIPOLAR FORCEPS INSTRUMENT | ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL, INC. | 420172-06 | 0710301 824 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ELECTROSURGICAL UNIT| DA VINCI S SURGICAL SYSTEM| ACCESSORIES |