Report Number
3004209178-2008-03816
Event Type
Injury
Date Received
July 8, 2008
Date of Event
April 1, 2008
Report Date
June 11, 2008
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
LGW
PMA / PMN Number
P840001
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

WHEN THE PT'S CAMPING TRAILER WAS PLUGGED INTO AN ELECTRICAL OUTLET, THE PT EXPERIENCED A SHOCKING SENSATION WHENEVER HE TOUCHED THE METAL OF THE CAMPER WITH HIS STIMULATOR TURNED ON; HE DID NOT FEEL IT WHEN THE STIMULATOR WAS TURNED OFF. WHEN THE CAMPER WAS PLUGGED INTO A DIFFERENT OUTLET, IT STOPPED. HE DID RECEIVE SOME SMALL SHOCKS ON OCCASION WHEN HE TOUCHED DIFFERENT ELECTRICAL ITEMS; THE PT THOUGHT IT MUST BE WHEN SOMETHING WASN'T GROUNDED PROPERLY. HE SPOKE TO A CONTRACTOR WHO STATED MOST PEOPLE WOULDN'T HAVE FELT THE SHOCK; THE PT STATED THAT HE MUST BE MORE SENSITIVE NOW. NO ONE ELSE FELT THE SHOCKING SENSATION UNTIL THE PT TOUCHED THEM WHILE THEY WERE TOUCHING THE CAMPER. USING A VOLTAGE METER GROUNDED ON ONE SIDE AND THE OTHER IN HIS HAND (WHILE TOUCHING CAMPER) HE HAD A READING TOPPING OUT AT 17 VOLTS. THE PT'S STIMULATOR WAS LOCATED IN HIS BACK. THE PT WAS AT HOME. HIS STATUS WAS "GOOD". THE PT WAS ENCOURAGED TO CONTACT HIS HCP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RESTORE ADVANCED RECHARGEABLE LGW MEDTRONIC PUERTO RICO OPERATIONS CO. 37713 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention EXTENSION MODEL 37082 LOT# NKB011382N| EXPLANTED| IMPLANTED| IMPLANTED| LEAD MODEL LEAD LOT# UNK| RECHARGER MODEL 37752| PROGRAMMER MODEL 37742 LOT# NJD057251N| EXPLANTED