BD RAPID DETECTION OF SARS-COV-2 VERITOR

MDR report key: 10712752 · Received October 21, 2020

Report

Report Number
3006948883-2020-00587
Event Type
Malfunction
Date Received
October 21, 2020
Date of Event
October 10, 2020
Report Date
April 28, 2021
Manufacturer
BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD.
Product Code
QKP
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: BD HAS RECEIVED SEVERAL CUSTOMER COMPLAINTS FOR FALSE POSITIVE RESULTS, WHEN USING BD SARS-COV-2 REAGENTS FOR BD VERITOR¿ SYSTEM. THE CURRENT INVESTIGATION CONCERNS MULTIPLE LOTS OF BD SARS-COV-2 REAGENTS FOR BD VERITOR¿ SYSTEM. BD TAKES A SYSTEMATIC APPROACH TO INVESTIGATING FALSE POSITIVE COMPLAINTS THAT ARE RECEIVED. THIS APPROACH INVOLVES REVIEW OF MANUFACTURING BATCH HISTORY RECORDS FOR BATCH NUMBER 0221748, TESTING OF RETENTION SAMPLES OF BATCH NUMBER 0221748, AND TESTING OF CUSTOMER RETURNED SAMPLES IF APPLICABLE. THE INVESTIGATION DID NOT FIND A ROOT CAUSE FOR THE FALSE POSITIVE RESULTS THAT WAS OBSERVED. THE ROOT CAUSE IS UNDER INVESTIGATION AND WILL BE DOCUMENTED IN OUR QUALITY SYSTEM. BD CANNOT CONFIRM THE COMPLAINT BASED ON THE INVESTIGATION THAT WAS PERFORMED. A CORRECTIVE AND PREVENTIVE ACTION (CAPA#1878253) IS ALREADY INITIATED TO INVESTIGATE THE ROOT CAUSE AND SOME MITIGATION ACTIONS ARE ALREADY BEING ADDRESSED.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD RAPID DETECTION OF SARS-COV-2 VERITOR ASSAY 3 FALSE POSITIVE RESULTS WERE OBTAINED. A REPEAT TEST WAS PERFORMED USING A PCR TEST METHOD AND THE RESULT WAS NEGATIVE. THE CUSTOMER STATED THE PATIENTS TESTED WERE ASYMPTOMATIC. THIS TEST IS NOT INTENDED FOR USE ON ASYMPTOMATIC PATIENTS AND WAS THEREFORE USED OFF LABEL. THE FALSE POSITIVE RESULTS WERE NOT REPORTED. THERE WAS NO REPORT OF PATIENT IMPACT. EUA # (B)(4).

Additional Manufacturer Narrative · 1

EUA # (B)(4). A SAMPLE IS AVAILABLE FOR EVALUATION. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED WHILE USING BD RAPID DETECTION OF SARS-COV-2 VERITOR ASSAY 3 FALSE POSITIVE RESULTS WERE OBTAINED. A REPEAT TEST WAS PERFORMED USING A PCR TEST METHOD AND THE RESULT WAS NEGATIVE. THE CUSTOMER STATED THE PATIENTS TESTED WERE ASYMPTOMATIC. THIS TEST IS NOT INTENDED FOR USE ON ASYMPTOMATIC PATIENTS, AND WAS THEREFORE, USED OFF LABEL. THE FALSE POSITIVE RESULTS WERE NOT REPORTED. THERE WAS NO REPORT OF PATIENT IMPACT. EUA # (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1176272 BD RAPID DETECTION OF SARS-COV-2 VERITOR CORONAVIRUS ANTIGEN DETECTION SYSTEM QKP BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD. 0221748

Patients

Seq Age Sex Outcome Treatment
1