FDA Adverse Event Injury Summary report: N

PEAK FX KEY SCR COUPLING ROD

MDR report key: 1071251 · Received July 7, 2008

Report

Report Number
1818910-2008-02584
Event Type
Injury
Date Received
July 7, 2008
Date of Event
June 13, 2008
Report Date
June 13, 2008
Manufacturer
DEPUY ORTHOPAEDICS, INC.
Product Code
LXH
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EXAMINATION OF THE RETURNED INSTRUMENT CONFIRMED THE TIP BROKE OFF. A DEPUY WARSAW MATERIAL RESEARCH MANAGER EXAMINED THE FRACTURED SURFACE. THE FRACTURED SURFACE SHOWS A SLIGHT TORSIONAL FRACTURE, HOWEVER, THE PREVAILING FRACTURE MODE IS CANTILEVER. A SECONDARY CRACK IS SHOWN AT THE ROOT OF THE THREAD INDICATING A BENDING ACTION IN CANTILEVER. A SEARCH OF THE COMPLAINT DATABASE FOUND NO PRIOR REPORTS FOR THIS PRODUCT NUMBER. BASED ON THE INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. DEPUY WARSAW CONSIDERS THE INVESTIGATION CLOSED. SHOULD ANY ADD'L INFO BE RECEIVED TO CHANGE THE OUTCOME OF THE PERFORMED INVESTIGATION, THE COMPLAINT WILL BE RE-OPENED.

Description of Event or Problem · 1

THREADED INSERTER INSTRUMENT BROKE OFF INSIDE THE KEYED LOCKING DEVICE EXTENDING THE SURGICAL PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PEAK FX KEY SCR COUPLING ROD 87LXH LXH DEPUY ORTHOPAEDICS, INC. NA AG0407

Patients

Seq Age Sex Outcome Treatment
1 62 YR