CORODYN A
Report
- Report Number
- 2243801-2008-00008
- Event Type
- Other
- Date Received
- July 3, 2008
- Date of Event
- June 4, 2008
- Report Date
- July 2, 2008
- Manufacturer
- B. BRAUN MEDICAL, INC.
- Product Code
- DYG
- PMA / PMN Number
- K822806
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE ACTUAL DEVICE IN THE INCIDENT WAS RETURNED TO THE MFR TO BE EVALUATED. THE BALLOON APPEARED TO BE SLIGHTLY DISCOLORED ON VISUAL OBSERVATION. ON OBSERVATION UNDER THE MICROSCOPE, IT WAS FOUND THAT THERE WERE SEVERAL ALMOST PARALLEL LINES ON THE SURFACE OF THE BALLOON INDICATING THAT THE BALLOON MATERIAL WAS DEGRADED DUE TO SOME UNCONTROLLED EXTERNAL FACTOR. THE SURFACE DEGRADATION ON THE BALLOON DID NOT APPEAR TO BE ATTRIBUTED TO THE MANUFACTURING PROCESS OF THE PART. AN ATTEMPT WAS MADE TO INFLATE THE BALLOON VIA STANDARD INFLATION SYRINGE. THE BALLOON COULD NOT BE INFLATED. ON PLACING BALLOON UNDER WATER AND TRYING TO INFLATE, IT WAS FOUND THAT THERE WAS A LEAK IN THE BALLOON SOMEWHERE NEAR THE MIDPOINT BETWEEN DISTAL AND PROXIMAL GLUE BANDS I.E. IN THE BALLOON MATERIAL ITSELF. NO OTHER PHYSICAL TESTING WAS POSSIBLE ON THE RETURNED SAMPLE. THE HOUSE RETAIN SAMPLES WERE TESTED PER SPEC AND FOUND TO PASS ALL VISUAL AND PHYSICAL TESTING. FURTHER TESTING WAS PERFORMED WITH INCREASED INFLATION VOLUME AND CYCLES AND NO FAILURES WERE NOTED. THERE ARE NO OTHER COMPLAINTS OF THIS NATURE REPORTED AGAINST THIS PART. NO SPECIFIC CONCLUSIONS CAN BE DRAWN.
RUPTURE OF ANGIOGRAPHY CATHETER; CUSTOMER INFLATED BALLOON BY CO2 GAS WITH FORMULARY QUANTITY, AND THEN THEY CARRIED IT TILL RIGHT VENTRICLE. AT THAT TIME THE BALLOON RUPTURED. SO THEY CHANGED THE CATHETER AND THE OPERATION WAS FINISHED. NO PT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CORODYN A | ANGIOGRAPHIC BALLOON FLOATATION CATHETER | DYG | B. BRAUN MEDICAL, INC. | AN505UP | 60829858 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |