FDA Adverse Event
Malfunction
Summary report: N
GE OEC 9800
MDR report key: 1071171
·
Received July 10, 2008
Report
- Report Number
- 1720753-2008-23095
- Event Type
- Malfunction
- Date Received
- July 10, 2008
- Date of Event
- June 10, 2008
- Report Date
- July 1, 2008
- Manufacturer
- GE OEC MEDICAL SYSTEMS INC.
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE X-RAY CONTROLLER BOARD, BAKCPLANE, AND BATTERIES WERE REPLACED AND THE WIRING TO THE TEMPERATURE SENSOR WAS REPAIRED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED, AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE 9800 SYSTEM HAD AN X-RAY TEMPERATURE SENSOR FAILURE ERROR, PRECHARGE ERROR, AND CONTROL PANEL ERROR MESSAGES DISPLAYED. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GE OEC 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |