FDA Adverse Event Malfunction Summary report: N

GE OEC 9800

MDR report key: 1071155 · Received July 10, 2008

Report

Report Number
1720753-2008-23110
Event Type
Malfunction
Date Received
July 10, 2008
Date of Event
June 17, 2008
Report Date
July 1, 2008
Manufacturer
GE OEC MEDICAL SYSTEM INC.
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE PS3 POWER SUPPLY WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED, AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE VERTICAL LIFT COLUMN BUTTON ON THE 9800 SYSTEM WOULD NOT WORK. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GE OEC 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEM INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1